Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 78298

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 10, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Medtronic Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Medtronic CareLink iPro Therapy Management Software, MMT-7340 The CareLink iPro Therapy Management Software (model, MMT-7340) is intended to be used by a healthcare professional (HCP) with the iPro2 CGM System and provides upload support for the iPro2 Recorder. The CareLink iPro software is used to upload glucose sensor data stored on a recording device, capture BG meter readings through manual entry or wired-upload, retrospectively calibrate sensor data, and provide retrospective reports of continuous glucose information. The reports generated by the CareLink iPro software are intended for use by healthcare professionals to assist with diabetes clinical assessment, patient education, and therapy optimization.

Z-0670-2018
Recall number
Z-0670-2018
Initiated
October 10, 2017
Classification
Class II
Status
Terminated
Recalling firm
Medtronic Inc.
Quantity
488 uploads

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software using the new CareLink iPro Uploader feature may result in incorrect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software Manufacturing/Software Deployment

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Due to a time conversion error, data uploaded from the iPro(TM)2 recorder to the CareLink iPro software using the new CareLink iPro Uploader feature may result in incorrect reports. This pertains only to reports generated between September 23, 2017 and September 27, 2017 that used the new CareLink iPro Uploader feature.

Code information

Version 1.12

Distribution pattern

Worldwide Distribution - US Distribution