device · product 1 of 2
LIFEPAK(R) defibrillation electrodes, Catalog # 11101-000016, MIN #3202380-006 The defibrillation electrodes are used only with LIFEPAK EXPRESS(R) AED, LIFEPAK CR(R) Plus AED, LIFEPAK 1000 defibrillator, or LIFEPAK 500 Biphasic AED s with a pink connector.
- Recall number
- Z-0186-2018
- Initiated
- October 27, 2017
- Classification
- Class I
- Status
- Terminated
- Recalling firm
- Physio-Control, Inc.
- Quantity
- 14,219 units in total
App-derived interpretation
The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols.
Official device-enrichment evidence · Sourced
Process control
Inspect official wording and provenance
Reason for recall
The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols.
Code information
713609, 717912, 713904, 718033, 715008, 719323
Distribution pattern
Worldwide distribution: US (nationwide) including Puerto Rico and Guam and countries of: Canada, Austria, Belgium, Czech Republic; Denmark, Finland, France, French Polynesia, Germany, Hungary, Ireland, Italy, Latvia, Netherlands, Australia, China, Hong Kong, Japan, Argentina, Mexico, Norway, Panama, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom, and New Zealand.