device · product 1 of 1
CP5 Centrifugal Pump, made up of the drive unit, the emergency system and the pump control panel. It used with an Electrical Remote-Controlled (ERC) tubing clamp and a bubble sensor.
- Recall number
- Z-0401-2018
- Initiated
- October 30, 2017
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Sorin Group Deutschland GmbH
- Quantity
- 1,318 units
App-derived interpretation
Possibility that the automatic line closure by the Electronic Remote-Controlled (ERC) clamp after detection of an air bubble could take longer than is intended by design in the CP5 system.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Possibility that the automatic line closure by the Electronic Remote-Controlled (ERC) clamp after detection of an air bubble could take longer than is intended by design in the CP5 system.
Code information
Model No. 60-00-60. Drive Unit: Ref #60-01-04, UDI 04033817900948. Emergency System: Ref #60-01-35, UDI 04033817900955. Pump Control Panel: Ref #60-02-60, UDI 04033817901006.
Distribution pattern
Distributed domestically to Distributed internationally to Algeria, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Canada, China, Colombia, Croatia, Czech Republic, Ecuador, Estonia, Finland, France, Germany, Hong Kong, Iraq, Italy, Japan, Jordan, Kuwait, Libya, Lithuania, Luxembourg, Malaysia, Morocco, Norway, Palestinian Territory, Panama, Peru, Poland, Qatar, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, The Netherlands, Tunisia, United Kingdom, Vietnam