device · product 1 of 1
Elekta ERGO++ Product Usage: ERGO is often used for stereotactic treatments.
- Recall number
- Z-0234-2018
- Initiated
- September 01, 2017
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Elekta Inc
- Quantity
- 182
App-derived interpretation
Incorrect DICOM mapping of the exported collimator or couch angles from ERGO, which would lead to incorrect rotation of the collimator or couch when using a MLC device for planning.
Official device-enrichment evidence · Sourced
Nonconforming Material/Component
Inspect official wording and provenance
Reason for recall
Incorrect DICOM mapping of the exported collimator or couch angles from ERGO, which would lead to incorrect rotation of the collimator or couch when using a MLC device for planning.
Code information
Version 1.7.3 and higher
Distribution pattern
Worldwide Distribution - US Nationwide in the states of: VA, CA, TX, MO, NY, WA, NM, MA, FL, CT, OK, SC, MS, NY, GA, NC and the countries of: AUSTRALIA, BANGLADESH, BRAZIL, BULGARIA, CANADA, COLOMBIA, CZECH REPUBLIC, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ITALY, JAPAN, JORDAN, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, PERU, PHILIPPINES, PORTUGAL, RUSSIA, SAUDI ARABIA, SINGAPORE, SOUTH KOREA, SPAIN, TAIWAN, THAILAND, TURKEY, VENEZUELA, and VIETNAM