MARS Treatment Kit, Use only with MARS(R) Monitor 1TC, GAMBRO, Product Code 800541
Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit.
These labels are deterministic app interpretations, not FDA categories.
Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit.
Code information
UDI: 37332414107728; Lot Numbers: 0000022597, 0000022648, 0000022663, 0000022671, 0000022721, 0000022767
Distribution pattern
US, Germany, Sweden, Italy, Belgium, Portugal, Spain, France, Russia, Poland, Turkey, China, Thailand, Taiwan, Hong Kong, Malaysia, Korea, and Mexico.