Event summary

Timeline bucket
October 09, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Cepheid

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Xpert CT/NG Catalog GXCT/NG-120 Batch 1000049568, Catalog GXCT/NG-10 Batch 1000053075, Catalog GXCT/NG-10 Batch 1000056156 and Catalog GXCT/NG-10 Batch 1000056157. Is a qualitative in vitro real-time PCR test.

Z-0592-2018
Recall number
Z-0592-2018
Initiated
October 09, 2017
Classification
Class II
Status
Terminated
Recalling firm
Cepheid
Quantity
3,586 kits (49,400 units)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Cepheid has received reports that some of our customers have experienced a higher number of temperature drift errors than would typically be expected when using this test and its intended use specimen types. Investigation of the issue also revealed an occasional probability of false negative results which remains consistent with package insert claims.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Cepheid has received reports that some of our customers have experienced a higher number of temperature drift errors than would typically be expected when using this test and its intended use specimen types. Investigation of the issue also revealed an occasional probability of false negative results which remains consistent with package insert claims.

Code information

Lot numbers: 25718, 25719, 25725, and 25727.

Distribution pattern

Worldwide Distribution -- USA, Puerto Rico, and Haiti.