Event 79078

4 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 04, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Fujifilm Medical Systems U.S.A., Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

4 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 4

Aspire HD (for FDR-1000AWS) The Fujifilm Digital Mammography System, Aspire HD (FDR MS-l000), generates full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.

Z-0659-2018
Recall number
Z-0659-2018
Initiated
August 04, 2017
Classification
Class II
Status
Terminated
Quantity
71 units (68 units Domestic. 3 units Foreign) in total

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and CR¿IR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.

Code information

1000AWS. Released Software Versions: V5.0, V5.1. Not Released Software Versions: V5.2, V6.0, V6.1, V7.0.

Distribution pattern

Worldwide Distribution.

device · product 2 of 4

Aspire HD Plus, Aspire HD-s (for FDR-2000AWS) The Fujifilm Digital Mammography Systems, Aspire HD Plus (FDR MS-2500) and Aspire ND-s (FDR MS-2000), generate full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and are intended for use in the same clinical applications as traditional screen-film mammography systems.

Z-0660-2018
Recall number
Z-0660-2018
Initiated
August 04, 2017
Classification
Class II
Status
Terminated
Quantity
71 units (68 units Domestic. 3 units Foreign) in total

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and CR¿IR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.

Code information

2000AWS. Released Software Versions: V5.1, V6.0, V6.1. Not Released Software Versions: V5.0, V5.2, V7.0.

Distribution pattern

Worldwide Distribution.

device · product 3 of 4

ASPIRE Cristalle (for FDR-3000AWS) The Fujifilm Digital Mammography System, ASPIRE Cristalle (FDR MS-3500) generates full field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.

Z-0661-2018
Recall number
Z-0661-2018
Initiated
August 04, 2017
Classification
Class II
Status
Terminated
Quantity
71 units (68 units Domestic. 3 units Foreign) in total

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and CR¿IR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.

Code information

3000AWS. Released Software Versions: V5.1, V5.2, V6.0, V6.1, V7.0. Not Released Software Versions: V5.0.

Distribution pattern

Worldwide Distribution.

device · product 4 of 4

Fuji Computed Radiography Mammography Suite, FCRMS (for CR-IR363AWS) The Fuji Computed Radiography Mammography Suite (FCRMS) is a software device that, in conjunction with a specified Fuji Computed Radiography system forms the Fuji Computed Radiography for mammography (FCRm) device. FCRm with a dedicated mammographic x-ray machine generates digital mammographic images that can be used for screening and diagnosis of breast cancer. It is intended for use in the same clinical applications as traditional screen-film based mammographic (SFM) systems. The mammographic images can be interpreted by a qualified physician using either hardcopy film or softcopy display at a workstation.

Z-0662-2018
Recall number
Z-0662-2018
Initiated
August 04, 2017
Classification
Class II
Status
Terminated
Quantity
71 units (68 units Domestic. 3 units Foreign) in total

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and CR¿IR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.

Code information

363AWS. Released Software Versions: V5.1, V6.0. Not Released Software Versions: V5.0, V5.2, V6.1, V7.0.

Distribution pattern

Worldwide Distribution.