Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 79107

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 31, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Oakworks Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Foot Controls impacted for the following devices: OAKWORKS¿ Medical CFPM 300 Imaging Table OAKWORKS¿ Medical CFPM 301 Imaging Table OAKWORKS¿ Medical CFPM 400 Imaging Table OAKWORKS¿ Medical CFPM 401 Imaging Table OAKWORKS¿ Medical CFPMB 301 Bariatric Imaging Table OAKWORKS¿ Medical CFUR 301 Urology Table OAKWORKS¿ Medical CFUR 401 Urology Table OAKWORKS¿ Medical CFLU 401 Lithotripsy/Urology Table

Z-0703-2018
Recall number
Z-0703-2018
Initiated
July 31, 2017
Classification
Class II
Status
Terminated
Recalling firm
Oakworks Inc
Quantity
1,104

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The foot control used on the OAKWORKS¿ Inc., CFPM300, CFPM301, CFPM400, CFPM401, CFPMB301, CFUR301, CFUR401 and CFLU401 Imaging Tables can become lodged under the base or under the column of the table.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The foot control used on the OAKWORKS¿ Inc., CFPM300, CFPM301, CFPM400, CFPM401, CFPMB301, CFUR301, CFUR401 and CFLU401 Imaging Tables can become lodged under the base or under the column of the table.

Code information

MODEL #(s): "CFPM 300 "CFPM 301 "CFPM 400 "CFPM 401 "CFPMB 301 "CFUR 301 "CFUR 401 "CFLU 401 CATALOG #(s): "75208-T01 and 75209-T01 "75211-T01 and 75212-T01 "75204-T01 and 75205-T01 "75206-T01 and 75207-T01 "75213-T01 "75214-T01 "75215-T01 "75216-T01

Distribution pattern

USA (nationwide) Distribution