Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 79148

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 11, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Medical Systems (Cleveland) Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

IQon Spectral CT with 4.7.2 software version.- Model 728332, a Philips Computed Tomography X-Ray System

Z-0841-2018
Recall number
Z-0841-2018
Initiated
January 11, 2018
Classification
Class II
Status
Terminated
Quantity
63

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
incorrect treatment; " Software

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

23 Software related issues for the IQon Spectral CT with 4.7.2 software version. These issues include: " Software errors that may result in CT spectral images misrepresentation or incorrect treatment; " Software errors that may result in partial set of images or inability to generate CT images.

Code information

Serial Numbers: 860008 860009 860010 860011 860012 860013 860014 860015 860016 860017 860018 860019 860020 860021 860022 860023 860024 860025 860026 860027 860028 860029 860030 860031 860032 860033 860034 860035 860036 860037 860038 860039 860040 860041 860042 860043 860046 860048 860049 860050 860051 860052 860053 860054 860055 860056 860057 860058 860059 860060 860061 860062 860063 860064 860065 860066 860067 860068 860070 860071 860072 890002 860047

Distribution pattern

Domestic and Foreign