device · product 1 of 2
ST AIA-PACK¿ PROG II, Progesterone II Assay, Part Number: 025239
- Recall number
- Z-0825-2018
- Initiated
- February 13, 2018
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Tosoh Bioscience Inc
- Quantity
- 5579
App-derived interpretation
Falsely elevated progesterone assay values will occur when using certain progesterone assay part numbers due to cross reactivity of DHEA-S, a metabolite of the DHEA (dehydroepiandrosterone) supplement.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Falsely elevated progesterone assay values will occur when using certain progesterone assay part numbers due to cross reactivity of DHEA-S, a metabolite of the DHEA (dehydroepiandrosterone) supplement.
Code information
All lots
Distribution pattern
Distributed in 39 states: AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI.