Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 79343

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 14, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Electronics North America Corp.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Philips HeartStart FRx AED.

Z-1913-2018
Recall number
Z-1913-2018
Initiated
March 14, 2018
Classification
Class II
Status
Terminated
Quantity
Total for FRx and HS1 - 660,519 defibrillators

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
An issue with one of the electric components (a resistor) could result in the device not delivering a shock when needed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An issue with one of the electric components (a resistor) could result in the device not delivering a shock when needed.

Code information

Manufactured between September 2002 and February 2013. The year of manufacture can be identified by the 2nd and 3rd characters in the serial number on the back of the AED in the following ranges: Home/Onsite: A02x-xxxxx through A13B-xxxxx; FRx: B04x-xxxxx through B13B-xxxxx NOTE: If the device was manufactured in 2013 and the 4th digit is the letter "C" or later (D, E, F...), it is not covered by the recall (i.e. A13G-02375). In some cases, a device within the date range covered by the notification did not contain a resistor previously associated with a failure. In that case, a notification was not sent. Confirmation as to whether the device contains the resistor at issue can be done by contacting the firm.

Distribution pattern

Distribution was nationwide and to Puerto Rico. There was also government/military distribution. Foreign distribution was made to the following countries: Canada, Afghanistan, Algeria, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Brunei Darussalam, Cambodia, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Fiji, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Malaysia, Malta, Mauritius, Mexico, Morocco, Namibia, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Niger, Norway, Oman, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirate, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia, and Zimbabwe.

device · product 2 of 2

Philips HeartStart (HS1) Onsite/Home AED

Z-1914-2018
Recall number
Z-1914-2018
Initiated
March 14, 2018
Classification
Class II
Status
Terminated
Quantity
Total for FRx and HS1 - 660,519 defibrillators

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
An issue with one of the electric components (a resistor) could result in the device not delivering a shock when needed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An issue with one of the electric components (a resistor) could result in the device not delivering a shock when needed.

Code information

Manufactured between September 2002 and February 2013. The year of manufacture can be identified by the 2nd and 3rd characters in the serial number on the back of the AED in the following ranges: Home/Onsite: A02x-xxxxx through A13B-xxxxx; FRx: B04x-xxxxx through B13B-xxxxx NOTE: If the device was manufactured in 2013 and the 4th digit is the letter "C" or later (D, E, F...), it is not covered by the recall (i.e. A13G-02375). In some cases, a device within the date range covered by the notification did not contain a resistor previously associated with a failure. In that case, a notification was not sent. Confirmation as to whether the device contains the resistor at issue can be done by contacting the firm.

Distribution pattern

Distribution was nationwide and to Puerto Rico. There was also government/military distribution. Foreign distribution was made to the following countries: Canada, Afghanistan, Algeria, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Brunei Darussalam, Cambodia, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Fiji, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Malaysia, Malta, Mauritius, Mexico, Morocco, Namibia, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Niger, Norway, Oman, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirate, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia, and Zimbabwe.