Event 79428

15 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 15, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Medical Solutions USA, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

15 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 15

SOMATOM Emotion 6 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Z-1445-2018
Recall number
Z-1445-2018
Initiated
February 15, 2018
Classification
Class II
Status
Terminated
Quantity
1301

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
To inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

To inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Code information

10046766

Distribution pattern

USA (nationally) Distribution.

device · product 2 of 15

SOMATOM Emotion 16 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Z-1446-2018
Recall number
Z-1446-2018
Initiated
February 15, 2018
Classification
Class II
Status
Terminated
Quantity
1301

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Code information

10046789

Distribution pattern

USA (nationally) Distribution.

device · product 3 of 15

SOMATOM Perspective Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Z-1447-2018
Recall number
Z-1447-2018
Initiated
February 15, 2018
Classification
Class II
Status
Terminated
Quantity
1301

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Code information

10046733

Distribution pattern

USA (nationally) Distribution.

device · product 4 of 15

SOMATOM Perspective 16 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Z-1448-2018
Recall number
Z-1448-2018
Initiated
February 15, 2018
Classification
Class II
Status
Terminated
Quantity
1301

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Code information

10046734

Distribution pattern

USA (nationally) Distribution.

device · product 5 of 15

SOMATOM Scope Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Z-1449-2018
Recall number
Z-1449-2018
Initiated
February 15, 2018
Classification
Class II
Status
Terminated
Quantity
1301

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Code information

10046799

Distribution pattern

USA (nationally) Distribution.

device · product 6 of 15

SOMATOM Scope Power Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Z-1450-2018
Recall number
Z-1450-2018
Initiated
February 15, 2018
Classification
Class II
Status
Terminated
Quantity
1301

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Code information

(Not Supplied)

Distribution pattern

USA (nationally) Distribution.

device · product 7 of 15

SOMATOM Spirit Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Z-1451-2018
Recall number
Z-1451-2018
Initiated
February 15, 2018
Classification
Class II
Status
Terminated
Quantity
1301

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Code information

(Not supplied)

Distribution pattern

USA (nationally) Distribution.

device · product 8 of 15

SOMATOM Emotion (2003) Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Z-1452-2018
Recall number
Z-1452-2018
Initiated
February 15, 2018
Classification
Class II
Status
Terminated
Quantity
1301

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Code information

3804890

Distribution pattern

USA (nationally) Distribution.

device · product 9 of 15

SOMATOM Sensation 10 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Z-1453-2018
Recall number
Z-1453-2018
Initiated
February 15, 2018
Classification
Class II
Status
Terminated
Quantity
1301

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Code information

7543015

Distribution pattern

USA (nationally) Distribution.

device · product 10 of 15

SOMATOM Sensation 16 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Z-1454-2018
Recall number
Z-1454-2018
Initiated
February 15, 2018
Classification
Class II
Status
Terminated
Quantity
1301

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Code information

7393114

Distribution pattern

USA (nationally) Distribution.

device · product 11 of 15

SOMATOM Sensation Cardiac Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Z-1455-2018
Recall number
Z-1455-2018
Initiated
February 15, 2018
Classification
Class II
Status
Terminated
Quantity
1301

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Code information

7543106

Distribution pattern

USA (nationally) Distribution.

device · product 12 of 15

SOMATOM Sensation 64 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Z-1456-2018
Recall number
Z-1456-2018
Initiated
February 15, 2018
Classification
Class II
Status
Terminated
Quantity
1301

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Code information

8377520

Distribution pattern

USA (nationally) Distribution.

device · product 13 of 15

SOMATOM Sensation Open Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Z-1457-2018
Recall number
Z-1457-2018
Initiated
February 15, 2018
Classification
Class II
Status
Terminated
Quantity
1301

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Code information

8872017

Distribution pattern

USA (nationally) Distribution.

device · product 14 of 15

SOMATOM Sensation 40 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Z-1458-2018
Recall number
Z-1458-2018
Initiated
February 15, 2018
Classification
Class II
Status
Terminated
Quantity
1301

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Code information

8874427

Distribution pattern

USA (nationally) Distribution.

device · product 15 of 15

SOMATOM Emotion Duo (2003) Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Z-1459-2018
Recall number
Z-1459-2018
Initiated
February 15, 2018
Classification
Class II
Status
Terminated
Quantity
1301

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
To inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

To inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Code information

3813933

Distribution pattern

USA (nationally) Distribution.