Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 79558

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 18, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Angiodynamics Inc. (Navilyst Medical Inc.)

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Fluid Management Convenience Kits (Angioplasty Kits) intended to be used in Fluid Management and/or Invasive Pressure Monitoring systems.

Z-1390-2018
Recall number
Z-1390-2018
Initiated
January 18, 2018
Classification
Class II
Status
Terminated
Quantity
38 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
NAMIC Fluid Management Convenience Kits contains a Fluid Delivery Set that was not assembled in accordance with specification, and prevents the device from being used, resulting in a minor delay in procedure.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

NAMIC Fluid Management Convenience Kits contains a Fluid Delivery Set that was not assembled in accordance with specification, and prevents the device from being used, resulting in a minor delay in procedure.

Code information

Lot Numbers: 5284422 5286610 5283649 5290525 5286542

Distribution pattern

Recall conducted to end user level. Recall Notifications were delivered by Fed Express.