Exact FDA event identifier

Event 79575

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 02, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Beckman Coulter Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

BECKMAN COULTER iChemVELOCITY, Urine Chemistry System, Catalog Numbers: 800-3530, 800-7167, 800-7719, 800-7720, 800- 7106, 800-3049, 800-3050, 800-3079, 800-3080, 800-7190, 800-7713, 800-7714, 800-7715, 700- 7176-001, 700-7177-001 Product Usage: The iChemVELOCITY automated urine chemistry system is an in vitro diagnostic device used to automate the urine chemistry analysis profile using iChemVELOCITY Urine Chemistry Strips. The iChemVELOCITY can be used as a stand-alone system, as well as in an iQ200 Series system, a configuration given the proprietary name iRICELL as it is designed to be hardware and software compatible with iQ200 Series systems. It produces quantitative results for specific gravity, semi- quantitative results for glucose, blood, leukocyte esterade, bilirubin, urobilinogen, pH, protein, ketones, and ascorbic acid; and qualitative results for nitrites, color, and clarity. iChemVELOCITY strips are intended for use only with the iChemVELOCITY analyzer. In particular they are not intended for visual reading. The iChemVELOCITY is not intended to be used as a Point of Care (POC) analyzer. The measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections, and liver finction. Tests performed using the iChemVELOCITY are intended for clinical laboratory use and in vitro diagnostic use only.

Z-1643-2018
Recall number
Z-1643-2018
Initiated
March 02, 2018
Classification
Class II
Status
Terminated
Recalling firm
Beckman Coulter Inc.
Quantity
2225 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Beckman Coulter has determined that there is a potential for incorrect settings to be installed on North American iChemVELOCITY Urine Chemistry Analyzers. 1. The installation of International settings can result in the generation of erroneous, false low results for some of the analytes at some concentrations. 2. The installation of colors different from those stated in the Instructions for Use (Colorless, Straw, Yellow, Amber, Red, Blue) for output settings will result in incorrect reporting of colors. For example, if Green is the color choice for the output setting instead of Blue, Green will be reported.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Beckman Coulter has determined that there is a potential for incorrect settings to be installed on North American iChemVELOCITY Urine Chemistry Analyzers. 1. The installation of International settings can result in the generation of erroneous, false low results for some of the analytes at some concentrations. 2. The installation of colors different from those stated in the Instructions for Use (Colorless, Straw, Yellow, Amber, Red, Blue) for output settings will result in incorrect reporting of colors. For example, if Green is the color choice for the output setting instead of Blue, Green will be reported.

Code information

All Fielded Serial Numbers with North American Configuration settings

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of Brazil, Canada, Malaysia, Singapore, Taiwan, Thailand, Turkey