Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 79740

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 20, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Medical Solutions USA, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Syngo.plaza software VB10A model numbers: 10863171, 10863172, 10863173 Syngo.Plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific workflows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.

Z-1699-2018
Recall number
Z-1699-2018
Initiated
February 20, 2018
Classification
Class II
Status
Terminated
Quantity
92

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
Software upgrade to correct format of study dates and issues with Legacy Presentation States (annotations) in order to prevent

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Software upgrade to correct format of study dates and issues with Legacy Presentation States (annotations) in order to prevent potential patient misdiagnosis.

Code information

All Serial Numbers

Distribution pattern

USA (nationwide)

device · product 2 of 2

Syngo.plaza software Syngo.Plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific workflows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.

Z-1700-2018
Recall number
Z-1700-2018
Initiated
February 20, 2018
Classification
Class II
Status
Terminated
Quantity
92

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
Software upgrade to correct format of study dates and issues with Legacy Presentation States (annotations) in order to prevent

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Software upgrade to correct format of study dates and issues with Legacy Presentation States (annotations) in order to prevent potential patient misdiagnosis.

Code information

All Serial Numbers

Distribution pattern

USA (nationwide)