Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 79766

6 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 14, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Medical Solutions USA, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

6 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 6

MEVATRON M2 / PRIMUS Mid-Energy, Digital Linear Accelerator, Model No. 01940035 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.

Z-1397-2018
Recall number
Z-1397-2018
Initiated
February 14, 2018
Classification
Class II
Status
Terminated
Quantity
18

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

Code information

Serial no: 3383 5474 3296 3591 3618 3103 3986 3629 3136 70-4323 5092 3639 3274 2855 3644 4018 70-4347 3939

Distribution pattern

Worldwide Distribution- US Nationwide in the states: AL, CA, FL, GA, IL, IN, KY, MA, MI, MO, MS, NJ, NY, OH, OK, PA, RI, SC, TN, TX, WI, WV, and WY; and PR. International distribution to Nassau.

device · product 2 of 6

ONCOR Avant-garde, Digital Linear Accelerator, Model No. 05863472 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.

Z-1398-2018
Recall number
Z-1398-2018
Initiated
February 14, 2018
Classification
Class II
Status
Terminated
Quantity
1

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

Code information

Serial no: 3905

Distribution pattern

Worldwide Distribution- US Nationwide in the states: AL, CA, FL, GA, IL, IN, KY, MA, MI, MO, MS, NJ, NY, OH, OK, PA, RI, SC, TN, TX, WI, WV, and WY; and PR. International distribution to Nassau.

device · product 3 of 6

ONCOR Expression, Digital Linear Accelerator, Model No. 07360717 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.

Z-1399-2018
Recall number
Z-1399-2018
Initiated
February 14, 2018
Classification
Class II
Status
Terminated
Quantity
2

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

Code information

Serial no: 5245 5353

Distribution pattern

Worldwide Distribution- US Nationwide in the states: AL, CA, FL, GA, IL, IN, KY, MA, MI, MO, MS, NJ, NY, OH, OK, PA, RI, SC, TN, TX, WI, WV, and WY; and PR. International distribution to Nassau.

device · product 4 of 6

ONCOR Impression , Digital Linear Accelerator, Model No. 05857920 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.

Z-1400-2018
Recall number
Z-1400-2018
Initiated
February 14, 2018
Classification
Class II
Status
Terminated
Quantity
7

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

Code information

Serial no:70-4143 70-4296 70-4235 70-4344 4021 3878 5154

Distribution pattern

Worldwide Distribution- US Nationwide in the states: AL, CA, FL, GA, IL, IN, KY, MA, MI, MO, MS, NJ, NY, OH, OK, PA, RI, SC, TN, TX, WI, WV, and WY; and PR. International distribution to Nassau.

device · product 5 of 6

ONCOR Impression Plus, Digital Linear Accelerator, Model No. 05857912 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.

Z-1401-2018
Recall number
Z-1401-2018
Initiated
February 14, 2018
Classification
Class II
Status
Terminated
Quantity
15

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

Code information

Serial no: 3855 4019 4067 3953 5148 3793 5098 5222 3801 5388 3948 3755 5033 4015 5095

Distribution pattern

Worldwide Distribution- US Nationwide in the states: AL, CA, FL, GA, IL, IN, KY, MA, MI, MO, MS, NJ, NY, OH, OK, PA, RI, SC, TN, TX, WI, WV, and WY; and PR. International distribution to Nassau.

device · product 6 of 6

PRIMUS HI, Digital Linear Accelerator, Model No. 04504200 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.

Z-1402-2018
Recall number
Z-1402-2018
Initiated
February 14, 2018
Classification
Class II
Status
Terminated
Quantity
18

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

Code information

Serial no: 2958 3832 3241 3574 3519 3922 3924 3361 4036 3302 70-4185 3949 3547 3657 3622 3731 3565 5153

Distribution pattern

Worldwide Distribution- US Nationwide in the states: AL, CA, FL, GA, IL, IN, KY, MA, MI, MO, MS, NJ, NY, OH, OK, PA, RI, SC, TN, TX, WI, WV, and WY; and PR. International distribution to Nassau.