Event 79783

7 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 26, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Medical Systems Nederlands

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

7 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 7

MultiDim Eleva 708032 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.

Z-1603-2018
Recall number
Z-1603-2018
Initiated
March 26, 2018
Classification
Class II
Status
Terminated
Quantity
184

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

Code information

System Serial # 32 177 259 239 264 258 52 67 111 257 94 142 121 191 143 260 265 74 207 101 182 174 124 263 4 172 254 151 246 64 183 76 152 188 261 242 102 34 147 241 134 178 245 176 175 38 162 107 220 65 194 69 247 13 215 78 112 29 139 98 79 158 61 33 106 35 50 16 80 205 161 11 204 37 12 189 9 232 15 228 155 47 171 213 209 218 96 90 212 217 181 163 140 114 115 108 71 141 14 36 48 224 25 105 57 216 138 24 28 211 160 192 43 44 157 154 41 219 8 68 129 75 221 7 10 63 146 45 26 125 193 210 1 60 3 132 81 22 23 208 5 131 135 88 226 6 156 49 91 92 93 201 250 251 72 95 148 39 31 2 30 51 100 255 252 256 109 196 164 170 227 173 83 159 53 137 54 116 118 119 167 168 169 97

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.

device · product 2 of 7

708033 URODiagnost As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.

Z-1604-2018
Recall number
Z-1604-2018
Initiated
March 26, 2018
Classification
Class II
Status
Terminated
Quantity
20

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

Code information

System Serial # 8 14 1 64 15 4 3 7 18 48 16 2 59 60 17 11 13 21 5

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.

device · product 3 of 7

MultiDiagnost Eleva 708036 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.

Z-1605-2018
Recall number
Z-1605-2018
Initiated
March 26, 2018
Classification
Class II
Status
Terminated
Quantity
37

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

Code information

System Serial # 18 42 46 41 36 35 37 47 27 43 10 5 24 29 34 49 26 13 15 25 50 12 17 48 32 30 31 14 40 19 9 16 45 8 4 21 11

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.

device · product 4 of 7

Mutlidiagnost-Eleva with Flat Detector 708034 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.

Z-1606-2018
Recall number
Z-1606-2018
Initiated
March 26, 2018
Classification
Class II
Status
Terminated
Quantity
206

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

Code information

System Serial # 201 137 223 228 219 60 272 35 111 85 121 38 120 185 168 32 30 100 235 34 148 149 67 269 191 118 170 241 210 255 8 273 33 240 131 99 193 36 138 96 222 278 257 246 23 147 162 220 47 179 221 126 195 46 144 254 189 75 56 217 83 5 167 6 154 155 156 177 183 157 245 116 62 207 81 113 134 103 115 55 200 180 175 78 93 24 106 110 123 266 3 URF.173 14 202 208 247 225 28 216 262 9 176 261 259 174 114 68 63 122 135 205 224 39 98 80 22 251 142 253 84 181 264 163 73 94 95 160 199 206 159 112 196 151 25 238 150 4 117 50 58 51 79 249 119 204 133 18 13 15 16 17 129 244 227 234 130 31 12 27 11 128 171 237 76 1 143 248 109 42 43 44 172 213 212 2 48 127 19 82 69 70 65 242 54 194 136 190 152 105 233 71 97 132 197 53 211 72 187 209 41 146 125 268 26 77 40

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.

device · product 5 of 7

Mutlidiagnost-Eleva with Flat Detector Eleva with Flat Detector 708037 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.

Z-1607-2018
Recall number
Z-1607-2018
Initiated
March 26, 2018
Classification
Class II
Status
Terminated
Quantity
199

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

Code information

System Serial # 40 239 123 58 88 172 129 81 136 159 244 238 82 65 249 70 142 188 128 178 117 45 203 79 229 199 126 158 71 60 187 47 197 87 152 105 76 26 124 139 35 15 66 241 150 86 230 22 206 207 222 146 29 151 111 190 59 228 155 4 233 92 30 221 163 201 91 69 44 205 48 18 34 171 110 175 121 174 43 160 185 23 202 209 95 144 50 141 156 708032/14 72 93 98 193 262 189 118 218 153 33 104 53 216 177 116 38 64 196 49 181 37 180 164 240 135 183 17 125 170 54 NL73 113 90 24 6 27 42 73 246 133 84 214 83 192 149 169 62 61 191 195 100 227 236 97 99 101 102 46 219 103 226 173 89 140 2 36 20 80 32 161 5 51 94 12 176 212 179 25 96 39 154 63 182 157 85 200 234 19 132 21 165 13 120 41 10 56 55 3 7 8 106 114 108 213 14 115 107 31 112

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.

device · product 6 of 7

MultiDiagnostic-Eleva with Flat Detector 708038 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.

Z-1608-2018
Recall number
Z-1608-2018
Initiated
March 26, 2018
Classification
Class II
Status
Terminated
Quantity
184

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

Code information

System Serial # 40 239 123 58 88 172 129 81 136 159 244 238 82 65 249 70 142 188 128 178 117 45 203 79 229 199 126 158 71 60 187 47 197 87 152 105 76 26 124 139 35 15 66 241 150 86 230 22 206 207 222 146 29 151 111 190 59 228 155 4 233 92 30 221 163 201 91 69 44 205 48 18 34 171 110 175 121 174 43 160 185 23 202 209 95 144 50 141 156 708032/14 72 93 98 193 262 189 118 218 153 33 104 53 216 177 116 38 64 196 49 181 37 180 164 240 135 183 17 125 170 54 NL73 113 90 24 6 27 42 73 246 133 84 214 83 192 149 169 62 61 191 195 100 227 236 97 99 101 102 46 219 103 226 173 89 140 2 36 20 80 32 161 5 51 94 12 176 212 179 25 96 39 154 63 182 157 85 200 234 19 132 21 165 13 120 41 10 56 55 3 7 8 106 114 108 213 14 115 107 31 112

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.

device · product 7 of 7

MultiDiagnostic-Eleva with Flat Detector 708035 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.

Z-1609-2018
Recall number
Z-1609-2018
Initiated
March 26, 2018
Classification
Class II
Status
Terminated
Quantity
1

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

Code information

Pending

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.