Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 80050

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 02, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Smith & Nephew, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

smith&nephew ANTHOLOGY POROUS HIGH OFFSET STEM, 12/14 TAPER, 7H SIZE, 32 MM x 112 MM, REF 71356107 Hip prosthesis.

Z-1921-2018
Recall number
Z-1921-2018
Initiated
May 02, 2018
Classification
Class II
Status
Terminated
Recalling firm
Smith & Nephew, Inc.
Quantity
13 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm.

Code information

Lot Numbers: 17LM22019, 17LM10934, 17LM05900

Distribution pattern

US nationwide distribution.