Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 80069

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 22, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Valeris Medical, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Apollo XT Suture Anchor, 5.5 x 15 Product Usage: Apollo XT Suture Anchors are intended for: Shoulder - Rotator Cuff Repair - Bankart Repair - SLAP Lesion Repair - Biceps Tenodesis - Acromio-Clavicular Separation Repair - Deltoid Repair - Capisular Shift or Capsulolabral Reconstruction Foot/Ankle - Lateral Stabilization - Medial Stabilization - Achilles Tendon Repair Knee - Medial Collateral Ligament Repair - Lateral Collateral Ligament Repair - Posterior Oblique Ligament Repair - Illiotibial Band Tenodesis Elbow - Biceps Tendon Reattachment - Ulnar or Radial Collateral Ligament Reconstruction Hip - Capsular Repair - Acetabular Labral Repair

Z-2430-2018
Recall number
Z-2430-2018
Initiated
January 22, 2018
Classification
Class II
Status
Terminated
Recalling firm
Valeris Medical, LLC
Quantity
113 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The label on Apollo XT 5.5mm Suture Anchor (part # XTS-5515; lots 2358, 2362, and 2363) indicated that the anchor screw material was titanium, however that part number is made of PEEK.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The label on Apollo XT 5.5mm Suture Anchor (part # XTS-5515; lots 2358, 2362, and 2363) indicated that the anchor screw material was titanium, however that part number is made of PEEK.

Code information

Model No. XT-5515, Lot No. 2358, 2362, 2363

Distribution pattern

US Nationwide Distribution in the states of California and Oklahoma.