VITROS Chemistry Products PHYT Slides, Catalog Number 8298671. in vitro diagnostic
- Recall number
- Z-2344-2018
- Initiated
- May 09, 2018
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics
- Quantity
- 5712
App-derived interpretation
There is a potential for imprecise results when using the slides. Increase in imprecision when using PHYT coating 0165. This leads to higher and lower than expected results for both quality control samples and patient samples.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
There is a potential for imprecise results when using the slides. Increase in imprecision when using PHYT coating 0165. This leads to higher and lower than expected results for both quality control samples and patient samples.
Code information
Lot Numbers: 2616-0165-4866, 2616-0165-4904, 2616-0165-4870, 2616-0165-4869, 2616-0165-5266, 2616-0165-5307, 2616-0165-5233, 2616-0165-5232, 2616-0165-5611, 2616-0165-5610, 2616-0165-5648, 2616-0165-5649, 2616-0165-6218, 2616-0165-6275, 2616-0165-6220, 2616-0165-6221, 2616-0165-6219, 2616-0165-6276, 2616-0165-6791, 2616-0165-6832, 2616-0165-6792, 2616-0165-6787, 2616-0165-7225, 2616-0165-7226, 2616-0165-7223, 2616-0165-7214, 2616-0165-7222
Distribution pattern
Worldwide distribution. US nationwide, Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Russia, Singapore, Spain, and Sweden