Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 80220

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 09, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Konica Minolta Medical Imaging USA, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

CS-7, Model No. A4C1

Z-2281-2018
Recall number
Z-2281-2018
Initiated
May 09, 2018
Classification
Class II
Status
Terminated
Quantity
Unavailable

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Panasonic has issued a Mandatory Product Safety Notice, indicating a possibility of battery ignition (fire) in CF-C2 "Toughbook" laptops. Since Konica Minolta uses this laptop in some of its product configurations, it notified its customers to utilize a BIOS Utility for the laptops to minimize the risk of an incident while Panasonic develops a final solution.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Panasonic has issued a Mandatory Product Safety Notice, indicating a possibility of battery ignition (fire) in CF-C2 "Toughbook" laptops. Since Konica Minolta uses this laptop in some of its product configurations, it notified its customers to utilize a BIOS Utility for the laptops to minimize the risk of an incident while Panasonic develops a final solution.

Code information

All version numbers.

Distribution pattern

Nationwide distribution. Foreign distribution to Argentina, Brazil, Canada, Chile, Colombia, Ecuador, Japan, Mexico, Netherlands, and Peru.

device · product 2 of 2

ImagePilot, Model No. D9MA

Z-2282-2018
Recall number
Z-2282-2018
Initiated
May 09, 2018
Classification
Class II
Status
Terminated
Quantity
Unavailable

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Panasonic has issued a Mandatory Product Safety Notice, indicating a possibility of battery ignition (fire) in CF-C2 "Toughbook" laptops. Since Konica Minolta uses this laptop in some of its product configurations, it notified its customers to utilize a BIOS Utility for the laptops to minimize the risk of an incident while Panasonic develops a final solution.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Panasonic has issued a Mandatory Product Safety Notice, indicating a possibility of battery ignition (fire) in CF-C2 "Toughbook" laptops. Since Konica Minolta uses this laptop in some of its product configurations, it notified its customers to utilize a BIOS Utility for the laptops to minimize the risk of an incident while Panasonic develops a final solution.

Code information

All version numbers.

Distribution pattern

Nationwide distribution. Foreign distribution to Argentina, Brazil, Canada, Chile, Colombia, Ecuador, Japan, Mexico, Netherlands, and Peru.