Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 80403

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 02, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Bio-Rad Laboratories Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

PR4100 Microplate Reader US, Catalog Number 45591587 Product Usage: The PR4100 Microplate Reader is a 96 well absorbance reader for the measurement of light absorbance (optical density-OD) of liquid media. The instrument is intended to be used primarily for the in vitro diagnostic analysis of samples from the human body to obtain information oh physiological and pathological states.

Z-2541-2018
Recall number
Z-2541-2018
Initiated
April 02, 2018
Classification
Class II
Status
Terminated
Recalling firm
Bio-Rad Laboratories Inc
Quantity
153

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The connector of the power supply was capable of fitting other components provided with the system. However, the voltage output of the reader power supply is incompatible with the other component and, if incorrectly connected, could result in irreparable damage to the system.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The connector of the power supply was capable of fitting other components provided with the system. However, the voltage output of the reader power supply is incompatible with the other component and, if incorrectly connected, could result in irreparable damage to the system.

Code information

UDI: 00847865014151 Serial Numbers: 140600072, 160900991, 161001953, 1112001938, 1112001941 , 1112004075 , 1112004079 , 1112004083 , 1112004084 , 1112004086 , 1112004088 , 1112004093 , 1201003822 , 1201003826 , 1201003827 , 1207001007 , 1207001015 , 1207001017 , 1207001018 , 1208001583 , 1208001584 , 1208001586 , 1208001590 , 1208005236 , 1208005244 , 1208005245 , 1211000978 , 1211000983 , 1211006938 , 1301003205 , 1301003212 , 1303000299 , 1303000300 , 1303000302 , 1303000305 , 1303007688 , 1306003195 , 1307006044 , 1307006045 , 1307006054 , 1307006055 , 1308000967 , 1308000968 , 1308000969 , 1308000971 , 1308000973 , 1308003338 , 1308003341 , 1308003345 , 1308003346 , 1308006670 , 1311008542 , 1311008544 , 1311008546 , 1311008555 , 1311008557 , 1311008562 , 1311008564 , 1311008565 , 1401002580 , 1401002583 , 1401002587 , 1401007785 , 1401007786 , 1401007789 , 1401007792 , 1401007793 , 1404000065 , 1404000066 , 1404000067 , 1404000068 , 1404002677 , 1404002684 , 1404002685 , 1405005640 , 1405005641 , 1405005645 , 1405005648 , 1406000064 , 1406000065 , 1406000066 , 1406000067 , 1406000068 , 1406000070 , 1406000071 , 1406000074 , 1406006883 , 1406006886 , 1407004144 , 1407004145 , 1407004147 , 1407004149 , 1407004151 , 1407004152 , 1410010410 , 1410010415 , 1410010416 , 1411010876 , 1411010877 , 1411010880 , 1412001714 , 1412001716 , 1412001721 , 1501001523 , 1501001526 , 1501001526 , 1501005823 , 1501005824 , 1501005825 , 1501005828 , 1501005829 , 1501005832 , 1506008687 , 1506008690 , 1506008695 , 1510005389 , 1510005393 , 1510005394 , 1510005395 , 1510005397 , 1510005398 , 1510005399 , 1510005414 , 1601001552 , 1601001553 , 1601001559 , 1601001561 , 1601001562 , 1601001563 , 1603001786 , 1603005407 , 1605000088 , 1609000992 , 1609000993 , 1609000994 , 1609000995 , 1609000996 , 1609001301 , 1609001302 , 1609001303 , 1609001304 , 1609001305 , 1609001306 , 1609009231 , 1610001955 , 1701001785 , 1701001788 , 1701004204 , 1701004210 , 1710010533 , 1710010539 , 13060003204 , 16303005411 , s1411010876

Distribution pattern

US Nationwide Distribution inn the states of : AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OR, PA, SC, TN, TX, VA, VT, WA, WI, and Puerto Rico.