Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 80452

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 14, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Maquet SAS

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

VOLISTA StandOP Surgical Light, ARD568811961 VLT600DF AIM STP; ARD568821961 VLT600DF AIM STP QL; ARD568811911 VLT600SF AIM STP; ARD568821911 VLT600SF AIM STP QL; ARD568821910 VLT600SF STP QL; ARD568811901 VLT600SF AIM STP Product Usage: The Maquet VOLISTA surgical lights are designed to illuminate the body of a patient during surgical operations, diagnostics and treatment. StandOP VOLISTA is a system of ceiling-mounted surgical lights suited for installation in surgical suites, examining rooms, doctors surgeries and out-patient consultations.

Z-2502-2018
Recall number
Z-2502-2018
Initiated
June 14, 2018
Classification
Class II
Status
Terminated
Recalling firm
Maquet SAS
Quantity
151

App-derived interpretation

Manufacturing or process control reason.manufacturing_process_control · v1.0.0
Improper assembly

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Certain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that connects the light head to the axle fork could result in the light head bracket breaking, causing the light head to detach from the arm and become suspended only from the cabling inside the arm. Use of an affected device may result in serious injury to patient and/or medical staff if the light head detaches and drops, impacting patient and/or medical staff during demonstration, service, or use.

Code information

All Serial Numbers manufactured from March 10, 2014 until November 17, 2016 6.

Distribution pattern

Worldwide Distribution - US Nationwide in the states of AR, AZ, CA, CO, CT, FL, GA, IA, ID, KS, KY, LA, MD, ME, MO, NE, NY, OH, OR, TN, TX, UT, and WA. Foreign distribution to Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil Cameroon, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Ghana, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Korea, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Namibia, Netherlands, Nigeria, Oman, Pakistan, Panama, Paraguay, Poland, Portugal, South Korea, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Tanzania, Yemen, and Zimbabwe.