Event 80555

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 09, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
INO therapeutics LLC/dba ikaria

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Mallinckrodt Pharmaceuticals INOflo DS, NO Gas Control System. Product Usage; Device Classification: The INOmeter is an accessory to the nitric oxide delivery device which is used for the delivery of inhaled nitric oxide to patient. An INOmeter is affixed to each cylinder of drug product. Functionality of the INOmeter: The INOmeter is affixed on top of each INOflo cylinder valve in Japan. When the healthcare provider opens the INOmeter the compressed gas cylinder valve opens to begin flow of the nitric oxide in nitrogen compressed gas. The nitric oxide gas is administered to a patient via a nitric oxide delivery device. The amount of time that gas flow administered to a patient is captured in the INOmeter. Each INOmeter is programmed to include the drug product expiration date as listed on the INOflo drug product label. The INOmeter electronically communicates to the nitric oxide delivery device via infrared technology by way of the iButton and in so communicates the expiration date of the corresponding cylinder of INOflo drug product to the delivery device. The delivery device evaluates and compares the expiration date to the current date of therapy administration. If the drug product expiration date is prior to the current date, an error and alarm is registered by the nitric oxide delivery device and administration of the expired drug to the patient is stopped after 120 seconds. If this occurs, the healthcare provider can operate the nitric oxide delivery device manually and deliver the INOflo as indicated in the Instructions For Use (IFU). The INOflo delivery device has the functionality to administer INOflo in manual mode.

Z-2733-2018
Recall number
Z-2733-2018
Initiated
July 09, 2018
Classification
Class II
Status
Terminated
Quantity
108 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
INOmeters affixed to the INOflo cylinders in Lot 17FP-0570 were programmed with an incorrect expiration date (11/2017). For Lot 17FP-0570, the label affixed to each INOflo cylinder has the correct expiration date noted (11/2020).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

INOmeters affixed to the INOflo cylinders in Lot 17FP-0570 were programmed with an incorrect expiration date (11/2017). For Lot 17FP-0570, the label affixed to each INOflo cylinder has the correct expiration date noted (11/2020).

Code information

Drug Product Lot: 17FP-0570: Serial Numbers: QL15110251 QL10110347 QL10110129 QL14110221 QL15030431 QL16090656 QL10060287 QL10110899 QL13050382 QL14110580 QL15030433 QL16090682 QL11100366 QL10110640 QL13060172 QL14110183 QL15030239 QL16090726 QL10110167 QL11100211 QL11090274 QL14110364 QL15040218 QL16090216 QL11100120 QL10101263 QL11090302 QL14120583 QL15030268 QL16090578 QL08030908 QL10070478 QL11090309 QL14120592 QL15030287 QL16090742 QL10110845 QL11090396 QL13060801 QL15020176 QL15030302 QL16090839 QL10110487 QL10111032 QL13061068 QL14120190 QL15030305 QL17060229 QL10020777 QL10110878 QL13061305 QL14120207 QL15110453 QL17060244 QL10110981 QL11090364 QL13110288 QL14110239 QL15110515 QL17060339 QL10070487 QL11090353 QL13110607 QL14110273 QL15110565 QL17060408 QL10100716 QL04100498 QL14050878 QL14110348 QL15110566 QL17060433 QL10101152 QL10070427 QL14060106 QL15010810 QL16090250 QL17060447 QL10060228 QL10110293 QL14060244 QL15010831 QL16090265 QL17060453 QL08031575 QL10060326 QL10110842 QL15011726 QL16090266 QL17060522 QL10070686 QL10110319 QL13110100 QL15010742 QL16090439 QL17060525 QL10060315 QL10110682 QL14110251 QL15030223 QL16090650 QL17090297 QL10111117 QL13050250 QL14110539 QL15030248 QL16090654 QL17090334

Distribution pattern

In the country of JAPAN