device · product 1 of 1
MEVION S250/MEVION S250i Proton Beam Radiation therapy system Product Usage: Indicated for the delivery of radiation for the treatment of patients with localized tumors or other conditions susceptible to treatment by radiation.
- Recall number
- Z-2746-2018
- Initiated
- June 29, 2018
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Mevion Medical Systems, Inc.
- Quantity
- 7
App-derived interpretation
QFix has advised Mevion Medical Systems that they are identifying unauthorized combinations of QFix Couch Tops with non-QFix devices in clinical use. The kVue Oen Touch latch consists of a precision latch mechanism that can be damaged when used with incompatible (third-party) accessories. This can create unsafe conditions for use and may compromise the subsequent use of any kVue inserts, whether authorized or not. Use of unauthorized or unvalidated inserts may lead to an incorrect patient treatment or physical injury.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
QFix has advised Mevion Medical Systems that they are identifying unauthorized combinations of QFix Couch Tops with non-QFix devices in clinical use. The kVue Oen Touch latch consists of a precision latch mechanism that can be damaged when used with incompatible (third-party) accessories. This can create unsafe conditions for use and may compromise the subsequent use of any kVue inserts, whether authorized or not. Use of unauthorized or unvalidated inserts may lead to an incorrect patient treatment or physical injury.
Code information
S250-0001 through S250-0006 and S250i-0007
Distribution pattern
US Nationwide Distribution in the states of DC, FL, MO, NJ, OH, and OK.