Polymedco, Inc.,/Polymedco Cancer Diagnostic Products, Llc
Dossier provenance
Source snapshots represented here
openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 2
PATHFAST Myoglobin II reagent cartridge Catalog Number: PF1111-K PATHFAST¿ NTproBNP is a product for in-vitro diagnostic use with the in vitro diagnostic system PATHFAST¿ for the quantitative measurement of N-terminal-pro B-type Natriuretic Peptide (NT-proBNP) in heparinized or EDTA whole blood and plasma.
A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.
These labels are deterministic app interpretations, not FDA categories.
A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.
Code information
Lot T738 Exp: 6/30/2018 UDI: (01)04987595301230(17)180630(10)T738
Distribution pattern
US Distribution to states of: AR, CA, FL, ID, LA, MN, NY and TX.
device · product 2 of 2
PATHFAST NTproBNP Catalog Number: PF1061-KUS PATHFAST¿ NTproBNP is a product for in-vitro diagnostic use with the in vitro diagnostic system PATHFAST¿ for the quantitative measurement of N-terminal-pro B-type Natriuretic Peptide (NT-proBNP) in heparinized or EDTA whole blood and plasma.
A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.
These labels are deterministic app interpretations, not FDA categories.
A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.
Code information
Lot : A735 Exp: 10/31/2018 UDI: (01)04987595301636(17)181031(10)A735 and Lot: O736 Exp: 01/03/2019 UDI: (01)04987595301636(17)190131(10)O736
Distribution pattern
US Distribution to states of: AR, CA, FL, ID, LA, MN, NY and TX.