Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 80724

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 30, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Polymedco, Inc.,/Polymedco Cancer Diagnostic Products, Llc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

PATHFAST Myoglobin II reagent cartridge Catalog Number: PF1111-K PATHFAST¿ NTproBNP is a product for in-vitro diagnostic use with the in vitro diagnostic system PATHFAST¿ for the quantitative measurement of N-terminal-pro B-type Natriuretic Peptide (NT-proBNP) in heparinized or EDTA whole blood and plasma.

Z-3185-2018
Recall number
Z-3185-2018
Initiated
April 30, 2018
Classification
Class II
Status
Terminated
Quantity
272 KITS in total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.

Code information

Lot T738 Exp: 6/30/2018 UDI: (01)04987595301230(17)180630(10)T738

Distribution pattern

US Distribution to states of: AR, CA, FL, ID, LA, MN, NY and TX.

device · product 2 of 2

PATHFAST NTproBNP Catalog Number: PF1061-KUS PATHFAST¿ NTproBNP is a product for in-vitro diagnostic use with the in vitro diagnostic system PATHFAST¿ for the quantitative measurement of N-terminal-pro B-type Natriuretic Peptide (NT-proBNP) in heparinized or EDTA whole blood and plasma.

Z-3186-2018
Recall number
Z-3186-2018
Initiated
April 30, 2018
Classification
Class II
Status
Terminated
Quantity
272 KITS in total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.

Code information

Lot : A735 Exp: 10/31/2018 UDI: (01)04987595301636(17)181031(10)A735 and Lot: O736 Exp: 01/03/2019 UDI: (01)04987595301636(17)190131(10)O736

Distribution pattern

US Distribution to states of: AR, CA, FL, ID, LA, MN, NY and TX.