device · product 1 of 1
IM Abbott Alinity ci-series (abdicqii) accessory driver v8.00.0001 software product that can receive information from a hospital computer system (also called a Laboratory Information System [LIS]).
- Recall number
- Z-0203-2019
- Initiated
- August 29, 2018
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Data Innovations, LLC
- Quantity
- 9
App-derived interpretation
Quality control results are not processed for Abbott Diagnostics Alinity ci series (abdicqii) instrument. This has been determined to be a driver malfunction. When an instrument transmits results from a quality control (QC) sample ID ("Q^Control" in SPM segment), and the option Hold Results Until Complete is enabled, the results do not parse into IM. The communication Trace shows the sample being transmitted, but no information is logged in the Specimen Event Log (SEL). There is a potential for harm if QC results are not processed, downstream auto verification rules may misfire based on missing QC results leading to a delay in, or incorrect treatment. QC results will not be available for evaluation by the lab. There have been no reported instances of patient harm from this malfunction.
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
Quality control results are not processed for Abbott Diagnostics Alinity ci series (abdicqii) instrument. This has been determined to be a driver malfunction. When an instrument transmits results from a quality control (QC) sample ID ("Q^Control" in SPM segment), and the option Hold Results Until Complete is enabled, the results do not parse into IM. The communication Trace shows the sample being transmitted, but no information is logged in the Specimen Event Log (SEL). There is a potential for harm if QC results are not processed, downstream auto verification rules may misfire based on missing QC results leading to a delay in, or incorrect treatment. QC results will not be available for evaluation by the lab. There have been no reported instances of patient harm from this malfunction.
Code information
accessory driver version number 8.00.0001
Distribution pattern
Distributed to OK. Foreign distribution to Argentina, Hong Kong, Malaysia, Thailand, Germany, Jordan, Australia, and Switzerland.