Timeline bucket
September 24, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
PerkinElmer Health Sciences Canada, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

PerkinElmer QSight 210 MD Mass Spectrometer Simplicity 3Q MD v1.0 with Hot Fix 2 Software intended to identify compounds in human samples by ionizing the compound and separating the resulting ions. Part Number: BC004872

Z-0513-2019
Recall number
Z-0513-2019
Initiated
September 24, 2018
Classification
Class II
Status
Terminated
Quantity
18 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Under certain conditions when using PerkinElmer QSight 210 MD mass spectrometer with Simplicity 3Q MD and Hotfix 2 software, it is possible for the original data from the last acquired sample to be over-written when the automatic flush is enabled.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Under certain conditions when using PerkinElmer QSight 210 MD mass spectrometer with Simplicity 3Q MD and Hotfix 2 software, it is possible for the original data from the last acquired sample to be over-written when the automatic flush is enabled.

Code information

210MDBC709100008 210MD8C709140010 210MD8C709150009 210MDBC709160011 210MDBC709180012 210MDBC709200013 210MD8C710020014 210MDBC710050016 210MD8C710110018 210MDBC710170020 210MD8C710180022 210MD8C712220428 210MDBC801240030 210MDBC801250031 210MDBC802140032 210MD8C802150033 210MD8C802260035 210MDBC803190037

Distribution pattern