Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 81439

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 01, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Medical Components, Inc dba MedComp

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

C3 Wave App, v. 2.0.5

Z-0222-2020
Recall number
Z-0222-2020
Initiated
October 01, 2018
Classification
Class II
Status
Terminated
Quantity
24

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
When the iPad is updated with Apple iOS software version 12, the C3 application malfunctions. Upon launching a new procedure, the ECG waves do not appear on the screen, in surface or internal mode. If the C3 application is not operating properly, the PICC procedure must be completed without using alternate methods to confirm PICC tip placement.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software Manufacturing/Software Deployment

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

When the iPad is updated with Apple iOS software version 12, the C3 application malfunctions. Upon launching a new procedure, the ECG waves do not appear on the screen, in surface or internal mode. If the C3 application is not operating properly, the PICC procedure must be completed without using alternate methods to confirm PICC tip placement.

Code information

v. 2.0.5; all units

Distribution pattern

Distributed to accounts in AZ, CA, CO, FL, IL, LA, MA, ME, NC, OH, TN, VA, and WV. Foreign distribution to Canada, Croatia, Czech Republic, Great Britain, Greece, Italy, Netherlands, Slovenia, and Sweden.