Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 81562

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 22, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
ARJOHUNTLEIGH POLSKA Sp. z.o.o.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

ARJOHUNTLEIGH GETINGE GROUP Flites Passive Clip Sling - Product Usage: The Arjo Passive Clip Sling is a product intended for assisted transfer of patient/resident with limited ability to move. Designed to be used for one patient during hospital stay and assists the caregiver with everyday patient handling routines, such as bed and chair related transfers. REF MFA1000M-L-L1 RM0255_7

Z-0589-2019
Recall number
Z-0589-2019
Initiated
October 22, 2018
Classification
Class II
Status
Terminated
Quantity
10

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Inspection process at the manufacturer found nonwoven material was used in specific lots of product that may break wile in use during the patient/resident transfer. Patient/resident might fall on the floor. IN consequence, the patient might sustain a major trauma, head injury - damage to tissue that is irreversible or may lead to death.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Inspection process at the manufacturer found nonwoven material was used in specific lots of product that may break wile in use during the patient/resident transfer. Patient/resident might fall on the floor. IN consequence, the patient might sustain a major trauma, head injury - damage to tissue that is irreversible or may lead to death.

Code information

Model - MFA1000M-L-L1 Lot: DHD1301721 UPN 40679252100332

Distribution pattern