Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 81587

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 28, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Howmedica Osteonics Corp.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Tritanium Posterior Lumbar (PL) cage intervertebral body fusion implant. Intended for an open, less invasive, and minimally invasive PLIF or TLIF surgical procedure.

Z-0605-2019
Recall number
Z-0605-2019
Initiated
November 28, 2018
Classification
Class II
Status
Terminated
Recalling firm
Howmedica Osteonics Corp.
Quantity
28,461

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The surgical technique is being updated to caution against misuse due to reports of cage fractures occurring intra-operatively and post-operatively.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The surgical technique is being updated to caution against misuse due to reports of cage fractures occurring intra-operatively and post-operatively.

Code information

UDI: 07613327117943, All lots, Model Numbers 48950070, 48950076, 48950080, 48950086, 48950090,48950096, 48950100, 48950106, 48950110, 48950116,48950120, 48950126, 48950130, 48950136, 48950140, 48950146, 48951070, 48951076, 48951080, 48951086, 48951090,48951096, 48951100, 48951106, 48951110, 48951116, 48951120, 48951126,48951130, 48951136,48951140, 48951146, 48952070, 48952080, 48952090, 48952100, 48952110, 48952120, 48952130,48952140,48953070, 48953080, 48953090, 48953100, 48953110, 48953120, 48953130, 48953140, 48954076,48954086, 48954096,48954106,48954116, 48954126, 48954136, 48954146, 48955076, 48955086, 48955096, 48955106, 48955116, 48955126, 48955136, 48955146, Vl/48950090, Vl/48953090, Vl/48955096, 48950066, 48954066, 48951092, 48951112, 48951132, 48950092, 48950112, 48950132,48956086,48956096,48956106, 48956116, 48956126,48956136, 48956146,48956140,48957092,48957112, 48957132, 48956112, 48956132, 48952092,48952112, 48952132, 48953092, 48953112, 48953132, 48950102, 48950122, 48950142, 48951102, 48951122, 48951142, 48956076,48956092, 48956102, 48956122, 48956142, 48953102, 48953122, 48953142

Distribution pattern

US Nationwide