Captus 3000 Thyroid Uptake System, Model # 5430-0076 and 5430-0077
- Recall number
- Z-0550-2019
- Initiated
- November 05, 2018
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Capintec Inc
- Quantity
- 385
App-derived interpretation
There is a potential for the spring arm failure as a result of a broken tension rod (a component inside the arm). A break in the tension rod can casue the collimator to fall downward to its lowest point of travel, approximately 25 inches from the ground.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
There is a potential for the spring arm failure as a result of a broken tension rod (a component inside the arm). A break in the tension rod can casue the collimator to fall downward to its lowest point of travel, approximately 25 inches from the ground.
Code information
Serial Numbers 901100 through 901309. Serial Numbers 901310 through 901965 as of 1/15/19.
Distribution pattern
Worldwide distribution - US nationwide distributed in the states of AR, AZ, CA, CO, CT, DC, FL, IA, IN, LA, MD, MI, MN, NC, NJ, OH, OK, PA, SC, SD, TX, VA, WI, WV, and countries of Argentina, Brazil, Canada, China, Cyprus, Greece, Hong Kong, India, Japan, South Korea, Vietnam.