device · product 1 of 1
Guardian Connect App CSS7200 used on the iPhone, iPad, and iPod Touch devices.
- Recall number
- Z-1282-2019
- Initiated
- July 04, 2018
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Medtronic Inc.
- Quantity
- Approximately 9,328 registered users
App-derived interpretation
The application may be closed by the operating system without alerting the user the app is no longer running or communicating with the transmitter resulting in the user not receiving alerts that could be associated with hypoglycemic or hyperglycemic events.
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
The application may be closed by the operating system without alerting the user the app is no longer running or communicating with the transmitter resulting in the user not receiving alerts that could be associated with hypoglycemic or hyperglycemic events.
Code information
UDI - (01)00643169693142(10)
Distribution pattern
Distribution was only made OUS. Countries involved are: Austria, United Arab Emirate, Australia, Belgium, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Hungary, Ireland, Israel, Iceland, Italy, Republic of Korea, Luxembourg, Netherlands, Norway, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Turkey, and South Africa.