Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 81704

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 20, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Zimmer Biomet, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Biomet DVR Crosslock Screws- 2.7mm x 20mm, Sterile ITEM 131827120 Product Usage: The system is intended for stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion and reconstructions (osteotomies) of small bones of the hand, foot, radius, ulna, ankle, humerus, scapula and pelvis, particularly in osteopenic bone

Z-0708-2019
Recall number
Z-0708-2019
Initiated
November 20, 2018
Classification
Class II
Status
Terminated
Recalling firm
Zimmer Biomet, Inc.
Quantity
50 pieces

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
One lot of product labeled 2.7mm x18mm Lock Screw Square Drive (Lot 029970) contains 20mm locking screws and one lot of product labeled 2.7mm x 20mm Lock Screw Square Drive (Lot 029980) contains 18mm locking screws

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Packaging process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

One lot of product labeled 2.7mm x18mm Lock Screw Square Drive (Lot 029970) contains 20mm locking screws and one lot of product labeled 2.7mm x 20mm Lock Screw Square Drive (Lot 029980) contains 18mm locking screws

Code information

LOT 029980 UDI (01)00880304535589(17)280501(10)029980

Distribution pattern

Country of Japan

device · product 2 of 2

Biomet DVR Crosslock Screws- 2.7mm x18mm, Sterile ITEM 131827118 Product Usage: Usage: The system is intended for stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion and reconstructions (osteotomies) of small bones of the hand, foot, radius, ulna, ankle, humerus, scapula and pelvis, particularly in osteopenic bone

Z-0709-2019
Recall number
Z-0709-2019
Initiated
November 20, 2018
Classification
Class II
Status
Terminated
Recalling firm
Zimmer Biomet, Inc.
Quantity
50 pieces

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
One lot of product labeled 2.7mm x18mm Lock Screw Square Drive (Lot 029970) contains 20mm locking screws and one lot of product labeled 2.7mm x 20mm Lock Screw Square Drive (Lot 029980) contains 18mm locking screws

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Packaging process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

One lot of product labeled 2.7mm x18mm Lock Screw Square Drive (Lot 029970) contains 20mm locking screws and one lot of product labeled 2.7mm x 20mm Lock Screw Square Drive (Lot 029980) contains 18mm locking screws

Code information

LOT 029970 UDI (01)00880304535572(17)280501(10)029970

Distribution pattern

Country of Japan