Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 81886

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 04, 2019
Product types
Device
Classifications
Class II
Statuses
Completed
Recalling firm wording
Elekta Limited

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Elekta Unity

Z-0706-2019
Recall number
Z-0706-2019
Initiated
January 04, 2019
Classification
Class II
Status
Completed
Recalling firm
Elekta Limited
Quantity
2 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
In rare circumstances, due to reasons such as inadequate hospital facility maintenance, water leakage, and weakening of speaker fixings during the lifetime of the system, the speakers can become dislodged and be pulled into the bore.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

In rare circumstances, due to reasons such as inadequate hospital facility maintenance, water leakage, and weakening of speaker fixings during the lifetime of the system, the speakers can become dislodged and be pulled into the bore.

Code information

Serial Numbers: 1536549, 1536549

Distribution pattern

TX and WI, Germany, Denmark, Netherlands, United Kingdom