Event 81927

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 07, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Linet Spol. S.r.o.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

AVE 2 Birthing Bed

Z-0991-2019
Recall number
Z-0991-2019
Initiated
January 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Linet Spol. S.r.o.
Quantity
253 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
LINET has recently become aware of a potential safety issue for the AVE 2 Birthing Bed caused by the power cable breaking at the bed connection or the orange insulation of the power cable may fail causing a possible spark or possible exposure to a live conductor.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

LINET has recently become aware of a potential safety issue for the AVE 2 Birthing Bed caused by the power cable breaking at the bed connection or the orange insulation of the power cable may fail causing a possible spark or possible exposure to a live conductor.

Code information

Serial Numbers 4PPB0091 through 4PPB626

Distribution pattern

AL, OK, NJ, OR, WA, CA, MN, LA, AZ, IL, MO, CO, OH, KY, FL, KS, UT, TN, LA, DE, CT, TX