Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 82014

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 08, 2019
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Draegar Medical Systems, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Infinity Delta Family patient monitors The Infinity Delta Series (Delta/Delta XL/Kappa) monitors are intended to be used on adult, pediatric, and neonatal populations, with the exception of the parameter Cardiac Output, ST Segment Analysis, and arrhythmia which are intended for use in the adult and pediatric populations only; and tcp02, which for the neonatal population, is to only be used when the patient is not under gas anesthesia.-Omega and Omega S are intended to be used on adult, pediatric, and neonatal populations.

Z-0974-2019
Recall number
Z-0974-2019
Initiated
February 08, 2019
Classification
Class II
Status
Ongoing
Quantity
10699 in total

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
cybersecurity vulnerabilities

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The firm is recalling their Delta family of patient monitors software due to cybersecurity vulnerabilities, which may cause the device to reboot, lose functionality, and/or lose communication.

Code information

All software versions, all serial numbers.

Distribution pattern

US Nationwide distribution including Puerto Rico.

device · product 2 of 2

Omega Systems These devices are intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.

Z-0975-2019
Recall number
Z-0975-2019
Initiated
February 08, 2019
Classification
Class II
Status
Ongoing
Quantity
10699 in total

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
cybersecurity vulnerabilities

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The firm is recalling their Delta family of patient monitors software due to cybersecurity vulnerabilities, which may cause the device to reboot, lose functionality, and/or lose communication.

Code information

All software versions, all serial numbers.

Distribution pattern

US Nationwide distribution including Puerto Rico.