Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 82074

4 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 08, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
GETINGE US SALES LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

4 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 4

Maquet XS Flat Screen Monitor Holder, XS32 SPE, Configuration Part Number ARD22201040A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.

Z-0987-2019
Recall number
Z-0987-2019
Initiated
January 08, 2019
Classification
Class II
Status
Terminated
Recalling firm
GETINGE US SALES LLC
Quantity
6

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Other

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.

Code information

Serial Numbers 500009 500010 500011 500012 500013 500014

Distribution pattern

The products were distributed to the following US states: AZ and TX.

device · product 2 of 4

Maquet XS Flat Screen Monitor Holder, PWD50+DF XS32 K3 SPE, Configuration Part Number ARD268400140A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.

Z-0988-2019
Recall number
Z-0988-2019
Initiated
January 08, 2019
Classification
Class II
Status
Terminated
Recalling firm
GETINGE US SALES LLC
Quantity
6

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Other

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.

Code information

Serial Numbers 1 2 3 4 5 6

Distribution pattern

The products were distributed to the following US states: AZ and TX.

device · product 3 of 4

Maquet XS Flat Screen Monitor Holder, PWD70+DF HD R XS32 K3 SPE, Configuration Part Number ARD268400150A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.

Z-0989-2019
Recall number
Z-0989-2019
Initiated
January 08, 2019
Classification
Class II
Status
Terminated
Recalling firm
GETINGE US SALES LLC
Quantity
6

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Other

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.

Code information

Serial Numbers 500001 500002 500003 500004 500005 500006

Distribution pattern

The products were distributed to the following US states: AZ and TX.

device · product 4 of 4

Maquet XS Flat Screen Monitor Holder, SAT XS32 12 SPE, Configuration Part Number ARD2SAT00130A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.

Z-0990-2019
Recall number
Z-0990-2019
Initiated
January 08, 2019
Classification
Class II
Status
Terminated
Recalling firm
GETINGE US SALES LLC
Quantity
4

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Other

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.

Code information

Serial Numbers 500013 500014 500096 500097

Distribution pattern

The products were distributed to the following US states: AZ and TX.