Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 82329

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 30, 2018
Product types
Device
Classifications
Class I
Statuses
Terminated
Recalling firm wording
Beckman Coulter Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

UniCel DxH 800 Coulter Cellular Analysis System, REF numbers 629029, B24465, B24802, B68304, B66445, and B63322. Product Usage: The UniCel DxH 800 Analyzer is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The UniCel¿ DxH 800 Analyzer identifies and enumerates the parameters indicated below on the following sample types: Whole Blood (Venous and Capillary) o WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV, NE%, NE#, LY%, LY#, MO%, MO#, EO%, EO#, BA%, BA#, NRBC%, NRBC#, RET%, RET#, MRV, IRF Pre-Diluted Whole Blood (Venous and Capillary) o WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV Body Fluids (cerebrospinal, serous and synovial) o TNC and RBC

Z-1382-2019
Recall number
Z-1382-2019
Initiated
July 30, 2018
Classification
Class I
Status
Terminated
Recalling firm
Beckman Coulter Inc.
Quantity
5,414 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints received of sporadic erroneously elevated platelet results without flags or system messages. Thrombocytopenia may go unrecognized.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software Design Change

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints received of sporadic erroneously elevated platelet results without flags or system messages. Thrombocytopenia may go unrecognized.

Code information

All current software versions are impacted

Distribution pattern

Worldwide Distribution - US Nationwide including Puerto Rico. There was government and military distribution. Foreign distribution was made to Andorra, Australia, Austria, Belgium, Canada, China, Croatia, Egypt, France, Germany, Greece, Hungary, India, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lithuania, Mexico, Morocco, Myanmar, Netherlands Antilles, New Zealand, Portugal, Romania, Russia, Saudi Arabia, Serbia, Slovenia, Spain, Switzerland, Trinidad and Tobago, U.A.E, United Kingdom, and Vietnam.

device · product 2 of 3

UniCel DxH 600 Coulter Cellular Analysis System, REF number B23858. Product Usage: Note: The DxH 600 hematology analyzer is the same instrument as the DxH 800 (K140911), except that it is a benchtop unit with no instrument cart, computer stand, and no connectivity capabilities. DxH 800 intended use states: The UniCel DxH 800 Analyzer is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The UniCel¿ DxH 800 Analyzer identifies and enumerates the parameters indicated below on the following sample types: Whole Blood (Venous and Capillary) o WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV, NE%, NE#, LY%, LY#, MO%, MO#, EO%, EO#, BA%, BA#, NRBC%, NRBC#, RET%, RET#, MRV, IRF Pre-Diluted Whole Blood (Venous and Capillary) o WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV Body Fluids (cerebrospinal, serous and synovial) o TNC and RBC

Z-1383-2019
Recall number
Z-1383-2019
Initiated
July 30, 2018
Classification
Class I
Status
Terminated
Recalling firm
Beckman Coulter Inc.
Quantity
2,059 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints received of sporadic erroneously elevated platelet results without flags or system messages. Thrombocytopenia may go unrecognized.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software Design Change

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints received of sporadic erroneously elevated platelet results without flags or system messages. Thrombocytopenia may go unrecognized.

Code information

All current software versions are impacted

Distribution pattern

Worldwide Distribution - US Nationwide including Puerto Rico. There was government and military distribution. Foreign distribution was made to Andorra, Australia, Austria, Belgium, Canada, China, Croatia, Egypt, France, Germany, Greece, Hungary, India, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lithuania, Mexico, Morocco, Myanmar, Netherlands Antilles, New Zealand, Portugal, Romania, Russia, Saudi Arabia, Serbia, Slovenia, Spain, Switzerland, Trinidad and Tobago, U.A.E, United Kingdom, and Vietnam.

device · product 3 of 3

UniCel DxH 900 Coulter Cellular Analysis System, REF number C11478 Product Usage: Note: The DxH 900 hematology analyzer is CLIA categorized under document CR180294 and is currently marketed as a member of the instrument family of the DxH 800 cleared under K140911, per FDA s guidance, Guidance for Industry and FDA Staff, Replacement Reagent and Instrument Family Policy . DxH 800 intended use states: The UniCel DxH 800 Analyzer is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The UniCel¿ DxH 800 Analyzer identifies and enumerates the parameters indicated below on the following sample types: Whole Blood (Venous and Capillary) o WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV, NE%, NE#, LY%, LY#, MO%, MO#, EO%, EO#, BA%, BA#, NRBC%, NRBC#, RET%, RET#, MRV, IRF Pre-Diluted Whole Blood (Venous and Capillary) o WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV Body Fluids (cerebrospinal, serous and synovial) o TNC and RBC

Z-1384-2019
Recall number
Z-1384-2019
Initiated
July 30, 2018
Classification
Class I
Status
Terminated
Recalling firm
Beckman Coulter Inc.
Quantity
341

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints received of sporadic erroneously elevated platelet results without flags or system messages. Thrombocytopenia may go unrecognized.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software Design Change

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints received of sporadic erroneously elevated platelet results without flags or system messages. Thrombocytopenia may go unrecognized.

Code information

All current software versions are impacted

Distribution pattern

Worldwide Distribution - US Nationwide including Puerto Rico. There was government and military distribution. Foreign distribution was made to Andorra, Australia, Austria, Belgium, Canada, China, Croatia, Egypt, France, Germany, Greece, Hungary, India, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lithuania, Mexico, Morocco, Myanmar, Netherlands Antilles, New Zealand, Portugal, Romania, Russia, Saudi Arabia, Serbia, Slovenia, Spain, Switzerland, Trinidad and Tobago, U.A.E, United Kingdom, and Vietnam.