Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 82459

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 19, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Medical Systems Nederlands

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Foot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 Product The Allura Xper FD Series are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures.

Z-1140-2019
Recall number
Z-1140-2019
Initiated
February 19, 2019
Classification
Class II
Status
Terminated
Quantity
7209 total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.

Code information

Foot Switches: 452270000141 Footswitch CV 3p 4m; 452270000151 Footswitch MD 3p 6m; 452270000142 Footswitch CV 3p 4m; 452270000381 Footswitch CV 3p 8m; 452270000382 Footswitch CV 3p 8m; 459800076001 Footswitch CV 4p+2 4m; 459800076021 Footswitch CV 4p+2 8m; 459800415571 Wireless Footswitch 4p+2; 459800415581 Wireless footswitch set 4p+2; 459800415531 Wireless Footswitch 3P; 459800415561 Wireless footswitch set 3P.

Distribution pattern

US Nationwide Distribution in the states of CA, FL, MI, OR, PA, TX and WA.

device · product 2 of 2

Foot Switches used with the following systems: 722001 Allura Xper FD 10 C 722002 Allura Xper FD10 F 722003 Allura Xper FD10 722005 Allura Xper FD10/10 722006 Allura Xper FD20 722008 Allura Xper FD20 Biplane 722010 Allura Xper FD10 722011 Allura Xper FD10/10 722012 Allura Xper FD20 722013 Allura XPER FD20 BIPLANE 722014 Allura Xper FD10 OR Table 722015 Allura Xper FD20 OR Table 722019 Allura Xper FD10/10 OR Table 722020 Allura Xper FD20 Biplane OR Table 722022 Allura Xper FD10 OR Table 722023 Allura Xper FD20 OR Table 722024 Allura Xper FD10/10 OR Table 722025 Allura Xper FD20 Biplane OR Table 722026 Allura Xper FD10 722027 Allura Xper FD10/10 722028 Allura Xper FD20 722029 Allura Xper FD20/10 722031 Allura CV20 722033 Allura Xper FD10 OR Table 722034 Allura Xper FD10/10 OR Table 722035 Allura Xper FD20 OR Table 722036 Allura Xper FD20/10 OR Table 722038 Allura Xper FD20/20 722039 Allura Xper FD20/20 OR Table 722058 Allura Xper FD20/15 722059 Allura Xper FD20/15 OR Table 722123 Field extensions Xper cardio 722124 Field extensions Xper vascular 722126 SmartPath to upgrade Allura to AlluraClarity cardio 722127 SmartPath to upgrade Allura to AlluraClarity vascular 722133 Field ext. Xper cardio systems R7.6 722134 Field ext. Xper vascular systems 722400 Cardio Vascular-Allura Centron 722030 INTEGRIS CV 722043 INTEGRIS Allura 15-12 (mono) Product The Allura Xper FD Series are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures.

Z-1141-2019
Recall number
Z-1141-2019
Initiated
February 19, 2019
Classification
Class II
Status
Terminated
Quantity
7209 total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.

Code information

Foot Switches: 452270000141 Footswitch CV 3p 4m; 452270000151 Footswitch MD 3p 6m; 452270000142 Footswitch CV 3p 4m; 452270000381 Footswitch CV 3p 8m; 452270000382 Footswitch CV 3p 8m; 459800076001 Footswitch CV 4p+2 4m; 459800076021 Footswitch CV 4p+2 8m; 459800415571 Wireless Footswitch 4p+2; 459800415581 Wireless footswitch set 4p+2; 459800415531 Wireless Footswitch 3P; 459800415561 Wireless footswitch set 3P.

Distribution pattern

US Nationwide Distribution in the states of CA, FL, MI, OR, PA, TX and WA.