Exact FDA event identifier

Event 82753

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 22, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Neusoft Medical Systems Co., Ltd.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

NeuViz 64 Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Z-1279-2019
Recall number
Z-1279-2019
Initiated
October 22, 2018
Classification
Class II
Status
Terminated
Quantity
37

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
Software defect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.

Code information

N64E120022E; N64E130039E; N641130051 E; N64E130062E; N64E130063E; N64E130066E;1 N641140077E; N64E140116E; N641140117E; N641140118E; N641140119E; N641140120E; N64EN150043E; N64E150121E; N64E150125E; N641150129E; N641150131 E; N641150136E; N641N160066E; N641N160067E; N64EN160068E; N64EN160071 E; N64EN160082E; N64EN160085E; N64EN160086E; N641N1 70093E; N641N1701 05E; N641N1 701 16E; N641N170120E; N641N170127E; N641N170128E; N641N170129E; N641N180156E; N641N180157E; N641N180161E; N641N180162E; N641N180205E

Distribution pattern

US Nationwide Distribution.

device · product 2 of 3

NeuViz 128 Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Z-1280-2019
Recall number
Z-1280-2019
Initiated
October 22, 2018
Classification
Class II
Status
Terminated
Quantity
6

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
Software defect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.

Code information

Series numbers N1281160084E; N1281160085E; N1281170161 E; N1281170231E; N1281180242E; N1281180248E

Distribution pattern

US Nationwide Distribution.

device · product 3 of 3

NeuViz 16 Essence Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Z-1281-2019
Recall number
Z-1281-2019
Initiated
October 22, 2018
Classification
Class II
Status
Terminated
Quantity
2

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
Software defect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.

Code information

Series numbers N16E180037E; N16E180047E

Distribution pattern

US Nationwide Distribution.