Exact FDA event identifier

Event 83096

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 16, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
GE Healthcare, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

GE Senographe Pristina - Product Usage: Senographe Pristina generates digital mammographic images that can be used for screening and in the diagnosis of breast cancer. Senographe Pristina is intended to be used in the same clinical applications as traditional film-based mammographic systems.

Z-2131-2019
Recall number
Z-2131-2019
Initiated
May 16, 2019
Classification
Class II
Status
Terminated
Recalling firm
GE Healthcare, LLC
Quantity
82

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potential slippage of the biopsy positioner when the gantry is rotated during a biopsy procedure.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential slippage of the biopsy positioner when the gantry is rotated during a biopsy procedure.

Code information

Serial# 716784BU4, 716785BU1, 716786BU9, 716787BU7, 716788BU5, 718583BU8, 718584BU6, 718585BU3, 718586BU1, 718587BU9, 718588BU7, 718589BU5, 718591BU1, 718592BU9, 718593BU7, 718594BU5, 718595BU2, 718596BU0, 718597BU8, 718598BU6, 718599BU4 , 718600BU0, 718601BU8, 718602BU6, 722648BU3, 722649BU1, 722653BU3, 722654BU1, 722655BU8, 722656BU6, 722657BU4, 730124BU5, 730125BU2, 730126BU0, 730292BU0, 730293BU8, 730663BU2, 730664BU0, 730665BU7, 732049BU2, 732050BU0, 732303BU3, 732304BU1, 732305BU8, 732362BU9, 732363BU7, 732448BU6, 732449BU4, 732450BU2, 732727BU3, 732728BU1, 733031BU9, 733903BU9, 733904BU7, 734170BU4, 734206BU6, 734840BU2, 735432BU7, 735779BU1, 736006BU8, 736007BU6, 736008BU4, 736366BU6, 736588BU5, 738021BU5, 738356BU5, 738357BU3 , 738358BU1, 738557BU8, 738573BU5, 738586BU7, 738587BU5, 738712BU9, 738792BU1, 738794BU7, 738795BU4, 738864BU8, 739150BU1 and 739721BU9.

Distribution pattern

Worldwide distribution - US nationwide distribution in the states of AZ, CA, FL, GA, IL, MI, MN, NC, NJ, NY, PA, TX, and VT, and countries of Algeria, Australia, Belgium, Canada, Chile, Columbia, Czech Republic, Finland, France, Germany, Hungary, Iceland, Japan, Korea (Republic of), Kuwait, Lebanon, New Zealand, Norway, Poland, Saudi Arabia, Senegal, South Africa, Sri Lanka, Sweden, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam.

device · product 2 of 2

GE Pristina Serena - Product Usage: Pristina Serena is an optional accessory of Senographe Pristina intended to provide accurate location of lesions in the breast in three dimensions.

Z-2132-2019
Recall number
Z-2132-2019
Initiated
May 16, 2019
Classification
Class II
Status
Terminated
Recalling firm
GE Healthcare, LLC
Quantity
16

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potential slippage of the biopsy positioner when the gantry is rotated during a biopsy procedure.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential slippage of the biopsy positioner when the gantry is rotated during a biopsy procedure.

Code information

Serial# 738192BU4, 738574BU3, 734192BU8, 733706BU6, 718590BU3, 735558BU9, 738023BU1, 737211BU3, 739178BU2, 733600BU1, 738696BU4, 738793BU9, 738695BU6, 734169BU6, 733707BU4 and 732113BU6

Distribution pattern

Worldwide distribution - US nationwide distribution in the states of AZ, CA, FL, GA, IL, MI, MN, NC, NJ, NY, PA, TX, and VT, and countries of Algeria, Australia, Belgium, Canada, Chile, Columbia, Czech Republic, Finland, France, Germany, Hungary, Iceland, Japan, Korea (Republic of), Kuwait, Lebanon, New Zealand, Norway, Poland, Saudi Arabia, Senegal, South Africa, Sri Lanka, Sweden, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam.