Exact FDA event identifier

Event 83144

24 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 07, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Burlington Medical, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

24 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 24

Frontal Stretch Buckle Closure Apron, Product Number A-10, Part Numbers REA10A, REA10B, REA10C, REA10*F, REA10*M, UEA10*M Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2107-2019
Recall number
Z-2107-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
9

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 423604 423524 423537 423549 387906 423539 423577 444669 384383

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 2 of 24

Frontal Apron Velcro Closure, Product Number F-17, Part Numbers REF17A, REF17B, REF17C, REF17D, REF17*F, REF17*M, REF17XLF, REF17XLM, REF17XXLM, REF173XLM, UEF17A, UEF17M, UEF17*F, UEF17*M, UEF17XLF, UEF17XXLM Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2108-2019
Recall number
Z-2108-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
104

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 387753 418758 414016 391249 381983 418764 381922 382933 382905 418853 387833 387759 423728 387770 387822 418760 418772 418762 418763 381759 414197 387955 443477 391067 418826 391060 387776 387777 387821 387758 387780 387769 387754 387824 387825 387826 423927 391167 387790 387182 391129 418883 442325 387838 359614 423723 443487 385516 381988 440775 384395 381987 381989 387227 418861 391057 443280 443272 443297 443298 419544 418943 387980 387823 423993 387818 384356 387174 444307 403409 440808 403707 410471 403144 440783 440810 405598 391103 382954 423708 387768 385674 419916 418941 423715 403737 423733 382889 381921 382984 391061 414010 418858 443292 385515 385522 387167 387239 387982 387752 418833 418958 418807 423861

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 3 of 24

Frontal Support Belt Apron, Product Number F-15, Part Numbers REF15*F, REF15*M, REF15B, REF15XLM Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2109-2019
Recall number
Z-2109-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
6

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 418940 444702 387785 442470 443381 381937

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 4 of 24

Frontal Quick Drop OR Apron, Product Number 601, Part Numbers ECRE601A, ECRE601B, ECRE601C, ECRE601XLM, RE601*M, RE601C, RE601D, RE601XXLM Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2110-2019
Recall number
Z-2110-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
23

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 381792 385576 423909 381833 385669 385700 381810 423908 423907 423936 423938 423910 423939 423906 423913 385705 423627 385664 385671 379711 385670 423911 423915

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 5 of 24

1 pc Full Wrap Support Belt Apron, Product Number F111, Part Numbers MTREF111*F, MTREF111E, MTREF111F, REF111*F, REF111*M, REF111B, REF111E, REF111F, SEF111*F, SEF111*M, SEF111C, SEF111D, SEF111E, SEF111F, UEF111*F, UEF111*M, UEF111A, UEF111B, UEF111C, UEF111D, UEF111E, UEF111XLM, UEF111XXLM Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2111-2019
Recall number
Z-2111-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
51

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 379682 379535 423744 423745 387796 426462 419802 384492 377568 379681 379690 379698 386085 377724 382991 384347 385884 386083 381755 379715 385640 381868 426861 426802 426955 426895 426634 426809 426951 416567 426461 426789 426952 381869 443767 443802 377712 381865 377711 382966 381899 381866 382876 387805 382949 385982 423502 339345 423501 414165 414167

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 6 of 24

2 pc Full Wrap Vest & Kilt, Product Number VK, Part Numbers ECREVKA, ECREVKB, ECREVKC, ECREVKXLM, ECSEVKA, ECSEVKB, ECSEVKC, ECSEVKXLM, REVK*F, REVK*M, REVKB, REVKC, REVKE, REVKXXLM, SEVK*F, SEVK*M, SEVKB, SEVKC, SEVKE, SEVKF, SEVKG, SEVKXLF, SEVKXXLM, UEVK*F, UEVK*M, UEVKXXLF Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2112-2019
Recall number
Z-2112-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
310

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 442293 443188 385619 387938 385611 385530 381778 387881 381782 387884 381781 387890 381843 387940 391124 387889 381780 387880 419929 418856 385607 387808 382887 391161 382951 381851 382896 385579 385748 381856 381857 381946 387866 423956 387877 423947 387872 423942 387871 423987 387863 423960 387869 424000 423961 387873 387864 419806 387870 419807 419776 423855 387874 423998 419774 419808 419770 419795 419775 419803 418868 419800 387867 419804 382803 391117 382956 382870 382941 382869 419942 419870 419943 418846 381831 385673 1111384500 382794 419934 418859 385610 385535 385735 385620 385740 385616 385533 385624 418779 387968 418789 418752 384333 382911 382878 382945 382856 382926 381903 381913 381864 391068 414763 418992 382959 382797 387924 418985 414044 423560 414346 414117 417567 419809 442291 443283 387865 418956 423601 384488 385626 385529 385724 387956 382962 379689 419937 418881 385697 418946 423979 391149 385634 381797 385636 387991 387987 387901 381794 387995 387957 418809 387960 418817 387898 418776 387977 418777 381804 418795 391156 418770 391151 418771 385637 387999 381904 385525 419847 387962 418895 387923 382985 382928 381850 381806 381830 381852 381872 391078 385633 385662 385622 385716 385627 385677 419944 419778 384477 384489 419913 423860 419945 419855 391125 381848 444520 443809 444290 443820 443801 443811 443344 426767 419938 423973 419940 418984 381846 387854 391131 385722 391157 382884 387970 388000 391106 385578 382892 385592 387142 387132 384496 387147 384475 384379 384381 384498 382942 384401 391127 385728 391105 382907 382979 382855 443307 443372 382987 384416 382836 382804 382965 391062 391120 382929 443402 443268 443220 443401 443817 443819 443315 442292 442290 443308 444717 444736 443346 427182 382970 387184 382977 382800 382910 382774 391065 382790 382795 382776 382939 382806 391058 382779 387950 419874 385621 385713 387868 418860 382971 384390 385730 391119 442294 443301 443302 442295 443371 443305 444718 443304 381906 381908 391122 387793 381907 391085 385629 385742 387992 391126 382927 382934 385609 387811 382904 385593 419797 423948 414766 418983 387976 419889 391137 387979 381814 385729 419786 419785

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 7 of 24

Vest, Single Item, Product Number Vest, Part Numbers ECREVESTD, ECREVESTF, ECREVESTXLF, REVEST*M, REVESTB, REVESTC, SEVEST*F, SEVEST*M, SEVESTA, SEVESTB, SEVESTC, UEVESTB, UEVESTXLM Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2113-2019
Recall number
Z-2113-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
22

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 382976 382964 385618 391130 391143 382858 381860 385617 381858 381854 387903 391141 442855 381789 387954 382861 391150 387944 387965 419791 385623 414166

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 8 of 24

2 pc Full Overlap Vest & Kilt, Single Item, Product Number FVK, Part Numbers REFVK*F, REFVK*M, REFVKB, REFVKD, REFVKE, REFVKF, SEFVK*F, SEFVK*M, SEFVKE, SEFVKXLM Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2114-2019
Recall number
Z-2114-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
129

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 382915 418869 443152 443303 443314 443116 443309 443122 443190 443115 385698 387994 387972 423969 381793 419868 385687 414095 419532 391154 423565 387997 385699 387917 382841 382946 382824 391134 384324 382787 381796 418843 387119 384454 384420 391099 382998 391096 382816 382909 423746 423667 382819 381798 384421 381821 384336 381828 385580 384349 423530 418903 423531 423976 419873 382859 418830 391146 419862 387967 418955 385691 391138 385680 385532 385692 387900 382782 384402 382823 382818 391163 382857 382866 382838 382906 381939 382995 381808 382903 391091 391164 381807 382862 418886 391148 418885 385690 391088 385681 391086 385682 391088 384412 381917 384422 384394 443131 443310 382900 385600 385693 391136 385683 391080 384294 384425 423945 387969 444689 443128 381790 418894 385686 391135 385684 385731 385689 381942 443104 444591 387974 419843 443129 443300 417565 391145 382850 382931

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 9 of 24

Overlap Vest, Single Item, Product Number FVest, Part Numbers REFVEST*F, REFVEST*M, REFVESTE, REFVESTXLM, SEFVEST*F, SEFVEST*M

Z-2115-2019
Recall number
Z-2115-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
7

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 385688 387953 385694 385696 387781 384424 382874

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 10 of 24

2 pc Full Wrap Vest & Kilt Back-Saver , Product Number PVK, Part Numbers REPVK*F, REPVK*M, REPVKE, REPVKF, REPVKXLM, SEPVK*F, SEPVK*M, SEPVKE, UEPVK*F Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2116-2019
Recall number
Z-2116-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
332

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 381795 385536 443407 444693 387809 385555 419989 419866 391144 385552 414184 419956 385597 381894 419988 419822 419992 419961 419912 423985 443459 444691 443411 443313 382943 384393 391038 379532 418751 385727 383000 382932 382986 382784 381893 385750 387236 382780 378516 381829 382875 391237 443398 443446 443476 444608 443410 443155 387927 385678 382812 382767 387915 418780 419828 419885 419869 423977 381901 385737 385550 385746 381891 385736 385547 387878 442492 443447 387934 387816 387973 387893 384413 391036 391101 382773 419990 418828 443471 444662 382957 384404 391087 385558 385562 385542 385512 385526 414081 414065 391224 382968 382763 382935 391228 382919 391227 382936 379707 377610 420000 419878 419994 419968 423553 419875 391229 381823 426470 443420 391044 382830 418914 418991 443312 443413 385741 381889 419927 387852 381883 381944 385596 381882 381900 385546 419962 423625 418898 418845 419987 423986 419991 419973 384415 382799 382832 382778 419983 423931 419964 387859 418773 418888 443425 444663 387933 387936 382913 382777 382845 382937 381888 385577 385523 385599 418949 419845 423965 418896 391225 387244 418871 423930 387810 385513 419976 423967 418876 418821 419864 423541 381890 385743 419824 418891 381895 382916 423903 419980 391049 384400 423515 419578 426456 429676 385511 391083 382752 382802 443799 443113 444538 443347 385549 385717 387858 387886 382994 382842 387930 418819 419856 418850 423631 414773 391063 391056 387882 418839 418890 379579 443181 443409 391039 384367 444724 443417 385733 381892 385559 385527 382975 382796 385563 385534 385952 384468 418783 418907 418912 418910 418837 418906 418913 418877 384455 382967 385594 391235 443470 444627 391053 382854 387853 418899 382980 382771 418884 423983 418852 419827 414324 414075 385598 382961 419999 418879 414062 414278 418887 387989 391050 382808 382825 382997 391055 382999 384396 391051 385551 418834 382791 384417 381896 381822 444544 426347 418902 418889 387928 387857 418987 419829 385554 387801 419984 423981 423978 419982 418857 419000 443282 443412 443416 443263 443260 443414 443359 443267 443343 429677 441453 441255 441465 441452 381845 385726 382783 391226 418866 418870 387896 418825 391040 382930 418849 418874 387899 418848 419985 423929 418862 418908 385548 387883 418841 423856 423966 418851 403119 443149

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 11 of 24

Peak Vest, Single Item, Product Number PVest, Part Numbers REPVEST*M, SEPVESTE, SEPVEST*F Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2117-2019
Recall number
Z-2117-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
7

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 418892 381887 381902 418811 418952 387916 429678

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 12 of 24

Kilt, Single Item, Product Number Kilt, Part Numbers ECREKILTF, ECREKILTXLF, REKILTA, REKILTB, REKILTXLF, REKILTXLM, REKILT*M, SEKILTB, SEKILTC, SEKILT*F, SEKILT*M Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2118-2019
Recall number
Z-2118-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
37

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 387197 384398 387814 418788 419871 387879 385725 385749 385732 387887 382947 419840 418873 423988 385738 387937 444696 418901 381799 419884 379705 386112 416433 415529 443261 418973 419967 419814 423980 419787 419892 391089 391128 385744 379537 417552 414129

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 13 of 24

Thyroid Collar, Magnetic Closure, Product Number TS5, Part Numbers RBTS5-R, RBTS5-T2T, RLTS5-R, RLTS5-T2T, SLTS5-R Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2119-2019
Recall number
Z-2119-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
210

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 419121 443916 408275 419300 419026 418172 418354 418174 418350 418362 418357 414627 418364 418363 418176 418358 418353 418352 418351 418349 418355 418359 418171 418175 418356 379202 418360 383132 414631 419298 408499 408495 429803 408423 408355 383130 424076 378575 378519 427547 419022 419028 419021 383160 414632 408413 408277 419024 383332 376923 378521 378520 383131 379205 419116 415973 409388 409389 408429 408418 422377 379201 379207 409387 409390 379199 414630 414629 418345 418361 418344 418438 418431 418177 443920 440828 433315 379161 379213 414834 441233 443685 407663 408415 408426 408425 408424 433368 441226 408496 418346 443682 433337 418407 383348 383347 414625 422460 414624 414635 383158 419379 419120 408416 418405 419377 383349 419375 383129 383126 419380 414831 414626 424075 383263 414634 383062 414633 379162 379227 379224 408278 441210 418410 383157 419299 408500 409998 418182 419378 388168 419025 420496 379203 379200 440827 441231 409999 383155 383156 383128 383127 419111 388169 378522 383133 383125 383350 419056 419027 433360 383161 414835 388171 440824 433311 441125 427541 443683 443674 441198 441055 415974 431461 408279 408276 418452 418448 441197 418180 424963 419374 433312 408422 379225 419118 419297 414832 408280 379206 441232 443676 418181 383162 414833 422374 418178 443675 441227 443677 443678 433316 443671 408383 418348 418347 383061 443684 419023 419376 419117 379226 379214 418450 440880 407618 378574 378577 378576 415000

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 14 of 24

1 pc Full Wrap Back Relief Apron, Product Number B52 Front & Back Apron, Part Numbers SEB52*F, UEB52*M Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2120-2019
Recall number
Z-2120-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
2

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 395561 395557

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 15 of 24

Thyroid Collar, Velcro Closure, Product Number TS2, Part Numbers RBTS2-R, RETS2-R, RLTS2PED, RLTS2-R, RLTS2-SEP, RLTS2-T2T, SBTS2-R, SETS2-R, SLTS2-R, UETS2-R, UETS2-SEP, ULTS2-R, ULTS2-SEP Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2121-2019
Recall number
Z-2121-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
2190

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 408295 388081 419292 419288 418098 418451 441111 441126 418276 414683 433375 418402 419210 376932 419356 424063 414941 418090 420827 420829 420824 418406 420882 422416 415855 408485 441114 441112 419019 378626 378642 419081 419342 419344 419345 406104 383177 383182 409423 408465 408432 408461 378560 414902 443732 433389 433390 433392 433326 433320 433324 433391 433377 433321 433318 443669 443728 443729 419064 418395 376802 378694 443934 440968 440887 388190 383019 383020 388027 409498 414539 414548 418201 418203 414938 420897 441203 414742 414686 414682 414681 414688 418295 386771 419253 420444 420446 420480 385447 419110 414981 419156 419115 419112 419157 385452 383166 383096 383097 383098 414748 388142 388146 388147 388150 417737 417736 418087 420316 409473 383106 383134 383135 418292 388059 422336 384664 384734 384733 384740 384679 383379 383152 383146 407620 408327 408328 428304 419338 388124 383150 387990 420289 383163 408388 408440 409495 414942 414943 420405 443738 414549 414553 378600 378603 378596 378606 378604 378598 378597 378599 378605 388039 418495 440830 414614 414615 414613 414620 414621 414619 414616 414622 414618 414617 440885 376801 376796 408382 407623 378639 424045 378264 378569 379216 378584 388051 384514 384513 378683 379217 384530 378682 378685 383011 379218 378678 378687 383009 384529 388049 378686 378679 378680 414808 440894 419012 419070 388035 378627 378635 378532 424006 385446 440853 414899 376858 419231 414879 383307 420305 414821 443878 441205 419399 455153 455155 433382 414906 419312 424072 408325 433352 441164 433385 419422 419418 385502 419421 418100 433269 441116 441190 441207 414798 388152 419096 419013 419089 440848 388189 419126 414501 383056 383058 383059 383057 441097 414998 419385 379231 378692 378700 383366 383356 383380 383370 441064 443957 441060 443962 414824 383406 414819 420363 376929 379238 379236 408352 408347 378585 378614 419425 408419 408372 378528 408350 441040 414696 415933 433384 414967 414986 414965 414964 414962 414963 414960 424860 408385 408386 408022 408023 408353 408406 408384 408403 408259 408407 408024 408028 408025 408408 408021 408269 408026 408009 408404 408387 408402 408263 408029 419187 424003 419337 419477 414684 443661 433439 443649 433428 440988 408875 407628 433265 429825 420439 384657 379028 379007 388126 376853 388129 388128 388040 384524 388026 418195 420838 418207 408006 420955 420252 420385 420384 406100 433460 417688 441138 441137 431729 383142 383143 418246 418293 408289 443620 384508 418434 418334 408343 408331 414976 408030 378564 376847 383110 388052 418238 424865 418229 379233 418341 443960 433327 433378 408375 384527 414778 378517 387795 419316 419247 419317 419315 391207 419322 419319 414749 418072 379699 388154 388050 414874 420956 419234 419233 378632 378641 379124 443942 441072 443955 443943 443954 427545 441206 440847 440840 379250 379243 384509 379221 379220 414849 383169 419321 414894 378527 378529 378537 378536 383281 383279 383004 388032 418335 422063 420784 418263 418387 418932 420806 376856 424029 414968 418321 420989 420991 409358 415944 409381 409437 409363 441036 441046 441208 420462 420770 420932 414885 414218 414538 414550 414536 424100 409475 433274 433272 409404 409410 378549 383343 419103 419320 423638 414958 414975 440993 379232 379247 433343 443722 420369 420371 418931 419011 378535 378628 424049 383280 383282 383040 378593 383050 383042 383100 419166 424025 414811 431687 414747 388175 388174 420312 441222 418400 418340 443877 414707 414710 381529 419266 384614 379126 388238 388246 378610 419075 419066 419071 378745 419077 419068 419054 378748 419082 378746 433261 376808 376797 419455 424103 424096 384535 419284 423656 423663 424110 424101 414994 388244 414641 383304 383303 376027 414654 420896 420905 414637 440971 383119 383117 383118 383112 383114 383107 383111 383113 383115 388236 388225 414944 433476 441122 406049 414535 443937 433404 384512 414510 414648 414645 419014 384528 408274 408356 388176 424030 419046 385649 376799 419314 419313 414744 417750 414745 417681 417731 417747 417708 417738 417745 424059 420946 417506 420953 420291 420387 420255 420310 443627 443720 443655 418449 418375 418378 418376 418436 418401 418437 418384 418374 418373 388242 420368 443734 408431 408453 419434 427549 441163 383153 376924 420301 408367 408333 408362 408291 408359 419248 378699 376920 378690 378695 378696 378697 378693 424008 419353 419350 419358 411696 378703 383179 378715 414913 378714 424074 378717 378716 388062 419114 418219 433379 441073 442096 433328 433325 418299 418298 418332 418285 418365 418324 418327 419178 383367 417740 414971 419038 419343 424004 414650 414653 414652 414651 418337 408008 383288 409432 440944 378741 424010 443736 419367 419484 419483 419479 419457 419352 419354 419456 419476 419481 422064 404192 404193 442257 442254 442259 388076 388043 388084 381624 408337 414687 418256 417744 418380 443745 419462 419460 433347 433346 424065 419052 419053 424026 388015 388066 388053 388055 388061 388070 419464 420306 419357 383138 383140 383121 383401 383355 408370 429911 408373 414636 424057 424050 424058 424061 378562 378561 440906 440907 440908 440902 440900 440905 443944 386764 419458 419310 408487 378551 383345 378548 383346 383344 378553 378550 378552 420785 420804 418926 386823 420805 409993 409995 440884 384525 424048 414817 414810 424005 414889 378590 378578 378589 378587 381528 384531 419360 383015 383014 419372 388047 383168 383164 381628 424018 408365 408368 408364 408369 440843 378518 417710 409969 418367 419149 419134 419129 419122 419142 419124 419125 419036 419130 388069 388065 388067 418486 420848 420833 420851 418487 418245 420854 418241 418490 414540 383013 414555 414509 414534 419826 387971 414556 414505 414703 419430 408376 420287 385453 419364 409364 433383 424046 378728 378729 407662 427542 379235 379242 379234 420261 420345 420346 420334 420337 420350 420342 420264 420336 420343 420348 420333 420352 420330 420344 420354 420332 420260 420358 420360 420341 420335 420347 420339 420353 420338 420340 420331 420349 420351 419395 419392 414995 414993 414992 420847 420834 420835 420836 420849 420842 420850 385451 383176 378702 419148 419351 414886 414887 414970 414953 425763 440892 440891 440893 440899 414895 414951 414812 414948 414946 414928 414876 419183 419431 418325 443746 443749 417702 409439 440897 441182 440896 440898 440895 419393 419402 419406 388247 404472 414537 414544 419291 409500 417727 417729 433329 433356 388014 419185 388063 388064 388068 388013 433283 441119 443933 443609 443602 443618 443607 433353 383183 383358 419006 383377 419007 419009 419015 419016 419008 419010 419018 419017 419048 409441 409472 384618 419084 383376 409468 408005 440831 378701 419473 419471 419465 419469 419475 419470 408438 408439 408488 419341 420293 419196 418326 420386 420314 420777 402841 408379 422268 419032 441000 414642 414991 378743 418342 420857 418488 418242 420981 419079 419091 419062 419072 419094 419073 419093 419090 419069 419063 419086 407649 409412 419311 424028 424111 419265 414644 414639 414647 379008 388028 418371 418333 418277 418377 441165 441156 443945 440882 443963 443948 404142 443956 441061 443952 443946 441062 444313 443941 440904 440881 440883 440888 440877 440890 443999 440876 440879 443998 440875 440878 444408 440869 441059 414815 408258 409984 383005 383002 378525 383030 378533 420282 420866 409374 440850 441211 441241 388024 385429 376921 378582 378591 384510 378580 376922 419087 419083 388153 378625 424023 386799 414507 414533 376877 422302 414542 382983 414508 418223 418192 443744 388060 383090 414898 414901 418366 418381 408433 419095 419065 419080 418288 383054 376957 424009 420771 414797 408380 441215 420283 420285 420284 414904 414596 424056 378681 409433 409408 424007 432458 379195 441081 414807 414700 378747 443615 443622 424044 424022 388240 383185 419347 419216 419348 419346 384515 383171 414939 414503 378594 420266 415857 440901 381632 381629 419035 419411 419423 419381 424051 414551 414541 414552 414545 414694 414746 408292 408334 419270 419209 419482 419480 408391 443887 433372 433373 383094 383093 383092 388235 383357 418231 418237 418221 418211 418220 418398 418386 418394 418389 418397 418370 418372 418391 418399 418396 431150 417735 418251 408354 408450 407624 414623 414649 386757 418188 383375 419005 383378 419003 419020 383374 419004 419002 383373 383372 417749 408491 443613 383287 441223 441250 414949 419227 419229 419390 414954 419388 419232 419226 414930 414888 419228 414893 414878 383181 383360 419230 419391 414945 414947 419398 419042 376859 383353 383361 424002 419193 419463 419459 417746 420909 420910 420912 420908 420911 414806 418209 420843 420845 440851 378725 419113 424020 414974 408366 443604 443612 443603 441194 388011 388080 441110 420767 409479 424079 424080 424093 424083 424081 424085 419255 419258 419254 419088 419259 419257 424104 424113 424106 424114 419263 419267 419252 424098 424112 419260 419078 419279 414751 414754 424087 414752 414753 424088 383141 424062 409471 441109 408267 408437 420265 420971 420965 420268 420267 420277 420978 420269 420272 420948 420278 420279 420270 420273 420274 420280 420276 420271 420963 420275 420952 420969 420966 420949 420967 420962 420974 420970 420968 420964 378684 408288 408282 408294 408284 408290 408271 383041 383052 441200 441155 419061 383351 420773 422300 379014 407654 418236 388078 419478 383363 418282 420899 381631 388041 414709 414705 414706 417684 414708 417682 418197 418222 378623 383189 383139 381669 408321 409478 414897 441145 440844 443884 441209 443597 440985 440994 440825 441117 440975 443598 440826 418311 418313 419092 378739 418210 418228 418193 418226 418216 378640 378631 408272 408268 409480 407644 409428 409421 409429 409418 441051 414891 440852 419416 376804 376789 419417 419415 419420 376807 419414 440834 409497 388117 376958 376960 388120 378689 409469 378545 414957 414936 388030 388249 424024 414883 419047 383137 407641 407634 431706 388222 409482 414638 384611 384596 384606 384602 388083 384597 384599 381661 376901 384598 381622 384595 381627 381630 433407 433376 408374 406084 406062 406091 406097 406092 420769 420903 440969 383302 378557 424881 408296 408287 381527 383154 419339 409411 408420 407636 407633 407619 407635 407632 407638 407627 407631 404453 384502 388243 383099 383188 383010 383012 414643 409417 409414 407625 443221 378698 409436 409425 409440 409416 409443 408261 408266 408262 408260 408265 408264 407621 424054 414937 414743 414905 388188 378558 383289 408361 419044 419242 378544 388125 376790 379168 431031 431688 443623 443614 443608 408338 417703 376881 388077 388023 383029 383024 383003 383035 383175 383172 378708 378705 378710 378707 383165 383173 383180 378713 378706 383174 378709 383178 378712 419152 409402 409438 409444 409397 409424 409405 409406 409392 409431 409399 409396 409422 409413 409395 409394 440832 384519 384516 384537 388226 384522 384520 384517 388223 388227 388224 414900 414822 419051 419251 419272 419276 419283 419278 419290 440941 388144 415992 407651 441124 433344 441089 443718 443719 443737 418243 388148 388186 441142 383049 441219 440987 440996 440997 440974 418187 418224 418230 418497 418196 418498 418194 418202 418190 418239 418189 418213 418235 418496 418199 418240 418232 418227 418215 418494 443593 433288 441175 441133 441178 427681 443594 433294 433291 433309 433293 441171 441174 441235 441234 433285 441159 441160 441162 441161 443631 443633 433286 433306 433310 427687 427683 388054 419218 419407 414597 419150 422467 424792 424866 424862 424870 424868 424861 424869 424864 409419 378742 378547 378563 383306 417748 417704 417707 441082 419468 419472 419474 376849 383053 383047 408007 408027 408482 408486 418338 418289 418383 418430 418388 418393 418392 418422 383277 419436 420317 440980 419181 419177 419168 419180 419179 419172 419169 419174 419182 419175 419184 419167 419170 419176 419163 419171 419159 419165 419173 419161 419235 419164 419158 387085 419160 419162 383089 383082 383087 383088 383091 419249 418309 418306 418301 418305 418316 422333 422332 420828 420762 418208 443747 443733 443653 424012 424019 424021 424015 424016 418088 419318 418247 420841 384536 420359 419404 386760 386759 379148 408401 443935 440992 414959 376794 422062 418250 441204 418261 418264 418262 419349 418320 418323 418322 418319 418079 418078 418076 418080 376792 376798 419131 376838 388195 383299 388197 388194 388177 383290 414820 408363 424067 424070 424069 424071 424068 424064 419467 424042 424039 419369 424041 419215 424040 424066 424043 419368 383405 424032 424031 424034 424037 423661 419195 419197 423662 424036 423657 423660 423658 424038 424035 424027 414977 414950 379010 383016 383017 383278 376854 418485 420256 420318 441144 427540 440841 440998 440978 440977 409442 443938 443936 440886 440967 440839 378609 419438 383409 383402 414704 418317 418429 379174 383170 419432 414892 414890 420307 409466 408878 418500 420794 420898 388025 414646 414546 414547 414803 420257 414795 419500 420263 419461 424014 418936 385448 424047 415807 416208 415829 422294 422274 422303 422304 422280 422271 422292 422278 417733 417730 417732 417734 418315 418314 378531 383371 418085 418089 418077 418084 418086 418483 418083 418081 418082 408848 408843 408839 424011 419400 419394 419439 424055 419441 419443 424053 418379 424052 419359 376795 433275 419382 419405 440951 440952 440955 440942 440953 423647 423652 423651 423653 423650 423646 423648 419429 406077 408479 376933 378613 378595 419280 418244 408447 408445 408442 408446 408452 408444 443619 443624 443797 431379 431722 419105 419104 378526 378530 414816 87354 378622 414880 378633 378568 441078 414952 418205 418233 418214 418185 418200 386817 418212 433345 443750 433395 433394 433397 433317 433322 443742 408392 418204 424117 424118 418492 418489 418493 409407 409426 388112 376850 383095 378611 383149 431047 431048 431666 419138 383297 388237 378523 383408 418491 415969 440943 443611 443605 433370 379146 388221 388183 381526 378691 376791 419355 376810 420906 409382 409373 440947 441134 440989 440949 414983 418339 418300 419186 378559 418294 420376 414931 376898 419363 379230 381525 384608 420774 383187 376959 383191 383144 417742 420259 385444 378619 388151 388149 418279 418259 418280 418284 418257 414333 414331 414330 418258 418260 408293 407648 419262 419268 419237 383404 419132 388057 422330 419362 423655 383365 383364 419203 441121 414969 420431 420435 420418 420420 420417 420429 420800 420430 420419 420422 420428 420424 420426 420421 420423 420427 420396 420958 420973 420951 420972 420959 420961 420957 420883 420943 420938 420940 420945 420947 420939 420944 420941 420942 420954 420960 420884 381635 419294 414809 414909 414932 406133 427538 433355 433357 419389 384511 418296 418297 418271 420788 419413 417694 431057 407622 381655 388019 388073 388012 408412 408409 408410 408411 407630 408358 418091 418096 386775 418278 386822 386821 418330 386820 386819 386818 418499 419396 383007 381658 383008 388022 419408 419410 424082 383407 419213 419214 414680 414933 414655 408498 440864 440866 440867 440868 440865 427692 427689 440903 444409 444407 441169 431679 431682 431680 431683 418101 407647 404089 423659 419397 418272 418265 418270 418266 383167 408251 408301 408302 408299 408297 408298 408300 409445 409376 388071 420309 418382

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 16 of 24

Thyroid Collar, Buckle Closure, Product Number TS4, Part Numbers RBTS4-R, RBTS4-T2T, RETS4-R, RLTS4-R, RLTS4-SEP, RLTS4-T2T, SBTS4-R, SETS4-T2T, SLTS4-R, UETS4-R Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2122-2019
Recall number
Z-2122-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
807

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 443646 419128 418103 418484 388010 376800 376806 419076 441180 441183 441186 441113 441181 441139 441136 441130 441131 441135 443748 423654 414987 420870 420877 443626 419123 388250 388245 383021 383032 381507 388123 381663 381508 376803 440957 440956 440966 433273 433267 440958 433270 414813 414814 414836 414837 414838 384534 388075 378637 414934 418310 418308 418302 418307 418303 383136 420902 424097 418102 418336 378749 409409 409435 381660 419139 383147 383145 419275 419274 418368 433348 408393 408010 408390 408316 409467 383081 388121 419246 424109 414884 408307 408322 408317 408313 408344 408330 408377 388248 378630 378638 378581 440846 433424 440842 441224 408308 408305 409448 408320 408310 408311 408323 408303 408312 408315 408306 408309 408340 408314 419043 383023 419074 431097 414823 424013 443695 443660 440979 443666 443651 443652 443648 443735 414896 419060 415987 415968 415965 415958 409372 415957 409430 409380 383259 414690 420357 378617 378618 383354 388056 378579 376925 388048 383359 384533 388048 376928 383184 419271 388034 419055 388009 414691 414697 414695 420900 420315 420933 385437 376926 376930 378616 440960 419198 419370 419426 417741 418254 419371 433400 433429 433438 441196 433427 441236 443665 433401 431374 443670 433441 433266 433500 429814 378586 384521 388074 422318 383048 420251 419194 378608 431162 443739 433485 418248 418249 383368 419401 420976 383044 383046 378612 419424 419412 414875 383051 422420 441039 414698 419269 419037 379180 419383 440845 417689 417690 441067 441069 443961 443953 441066 443958 441063 441068 414506 417709 417743 424060 415865 427558 440849 417695 419207 419264 384503 419335 419286 383291 383300 383305 383301 383273 383296 420852 420832 420853 420830 420855 418198 420846 420856 420840 420837 418225 420839 408493 408857 408405 408856 443725 419212 384610 419245 419433 388141 388184 388178 388241 388228 378538 383298 383266 383270 383294 383295 388018 388072 383293 383292 383256 383272 378541 378543 378542 378540 378539 418331 418369 384538 384539 418253 418252 414839 383074 383078 383075 383077 383080 383076 383079 383083 408319 383022 383026 414873 420977 420984 414693 376805 443724 420786 420775 420772 420765 420793 420760 420792 420797 420790 420791 420776 420778 420763 420768 420759 420766 420764 420779 420789 420787 388229 388230 388232 388233 376846 376844 384518 388231 414990 414989 414997 424852 414961 419221 424873 414877 419222 376867 376840 376839 376862 376842 376841 419135 429818 404041 409463 381569 419437 414640 414935 419200 419361 407626 414017 443637 443638 443636 443635 381570 378524 418092 440959 440946 420876 420286 420391 424099 419045 384592 383116 383086 419039 419277 419137 419224 420881 409403 441245 381666 381679 385454 408400 441127 379107 384617 379197 381633 419239 422437 388114 388127 379219 419133 419199 423649 420281 443964 443947 441185 441128 441129 441187 417705 417706 408394 407629 383412 414955 409983 408255 408256 409985 409992 408436 440871 420937 420304 414701 441217 440872 419208 419220 419223 420950 409434 414818 414842 418184 418433 418183 376809 408270 409477 414554 414956 388119 379110 379109 388118 417686 417683 417687 417685 418095 420904 420844 378567 383151 379172 440991 440970 409994 418287 409379 419136 383362 419154 419127 431498 424947 376857 388187 388198 383045 414988 419057 419282 417697 433478 378588 420761 422321 408492 407643 407645 408434 443723 376927 424084 433374 431123 442009 442010 431132 442008 431130 431122 431116 441150 441149 441143 431034 441153 441154 443875 443886 441148 441147 440984 440982 440976 440990 419427 383264 418385 424105 424108 420975 404422 441220 443664 388239 419205 440983 433271 424086 440945 443599 431704 414903 414929 409489 378565 388031 383275 383102 376851 383105 383104 383101 383103 383108 376855 383109 384506 378634 383084 383055 443610 431164 443625 376931 383276 414750 422317 409375 409483 419442 443721 433342 443704 433380 443700 443658 418283 414332 418286 417739 418291 418290 419041 419289 419040 419293 419050 419049 409486 441037 443876 441218 384656 419373 388166 388160 388161 388165 388162 388157 388164 388163 388115 378636 384526 419386 409362 409368 408395 443595 443596 388191 441213 442048 441189 409384 422428 427693 433406 433319 433295 433300 433301 433296 433297 433304 433303 433302 443634 433292 433307 441173 441170 441157 443629 433290 427688 443630 443628 443632 433289 441172 441158 441176 427684 433287 433308 441177 441179 427685 427680 384741 383410 414826 409420 378737 378738 419201 443879 418191 431134 414502 379182 420990 420988 419440 419387 419336 424107 419409 414881 418312 418094 378546 376843 388196 388193 419058 414973 419211 424102 418281 433495 408441 378704 419059 419202 440995 440950 378566 383085 419466 419261 419225 420880 420987 420783 414984 420414 414504 419958 414398 414400 419238 388033 384540 388159 388234 433444 443708 443716 441195 441166 433443 433475 441167 440981 443714 443641 388058 414999 414996 386783 419204 383018 419067 383369 418217 418234 418390 418206 441115 418099 404278 406082 406059 406094 406064 406065 406099 406105 406076 406068 406093 404280 406110 419366 419384 378573 419001 408389 443710 441152 427686 379018 383411 414511 388122 418186 418093 418329 420879 420901 414972 378711 383192 403470 408332 419281 419403 419217 419365 419287 404327 407660 407659 407661 407655 407658 407657 407656 408339 419428 418304 419241 419240 419244 419243 419206 388158 387002 381509 379115 414985 419219 418097 408466 386010 383260 389088 388021 388020 414966 408494 408443 407653 418927 408381 376852 414656 383190 408254 409447 409446 408348 408349 440873

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 17 of 24

Sleeve, Product Number Sleeve, Part Numbers RESL, SL 3 L, SL 3 M, SL 3 S, L 3 X, SL 3 XS, UESL Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2123-2019
Recall number
Z-2123-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
39

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 444749 418756 418757 408712 388138 388139 407957 339006 419340 391212 381909 381862 391205 414767 419574 419561 419562 419190 419189 391214 443199 418796 350481 407814 407813 443198 443200 443201 443197 391203 381803 379683 419191 419192 419107 383262 383261 387157 387158

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 18 of 24

Frontal Apron with Quick Drop Velcro Closure (no elastic), Product Number 401, Part Numbers RE401*M Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2124-2019
Recall number
Z-2124-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
1

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 418937

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 19 of 24

Front/Back Apron with Stretch Velcro Closure, Product Number EURO, Part Numbers ECSEEUROA, ECSEEUROB, ECSEEUROC, ECSEEUROXLM, REEURO*F, REEURO*M, REEUROC, REEUROE, SEEURO*F, SEEURO*M, UEEURO*F Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2125-2019
Recall number
Z-2125-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
48

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 381839 382008 381834 423928 423922 423917 423919 423921 423916 387073 423925 387072 423918 419789 387168 423923 423926 423924 387116 391152 381819 381824 391111 391093 387078 381786 382901 419907 419906 403296 382883 387075 391159 382894 391166 418938 387975 418911 382881 403349 405569 443366 377104 423704 403406 419550 419902 419903

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 20 of 24

Half Apron, Product Number Half Apron, Part Numbers REHALF*, REHALF3 Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2126-2019
Recall number
Z-2126-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
2

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 382955 418939

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 21 of 24

Cap, Product Number Cap, Part Numbers CAP1, CAP2 Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2127-2019
Recall number
Z-2127-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
49

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 389083 423885 413884 417509 384253 378507 378503 378505 418959 418960 423897 423899 423893 423895 414395 419567 417515 417514 420508 385648 414155 420509 415418 420507 420501 409316 409320 409330 403879 403880 403882 403881 409310 404000 441028 441030 441022 441031 441027 442251 443572 441025 441024 443579 443574 431256 443575 431517 443573

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 22 of 24

Hood, Product Number Hood, Part Numbers HOOD1, HOOD2 Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2128-2019
Recall number
Z-2128-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
3

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 419558 419557 414132

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 23 of 24

Porta Shield, Product Number Porta Shield, Part Numbers PTS2, PTS3, PTS4 Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2129-2019
Recall number
Z-2129-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
9

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 419886 443251 426551 443014 444704 416997 391132 419882 419877

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

device · product 24 of 24

Thyroid Collar Oversize, Product Number TS3, Part Numbers RBTS3-R, RETS3-R, RLTS3-R, RLTS3-SEP, RLTS3-T2T Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Z-2130-2019
Recall number
Z-2130-2019
Initiated
June 07, 2019
Classification
Class II
Status
Terminated
Recalling firm
Burlington Medical, LLC
Quantity
23

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Code information

Serial Numbers: 383071 418435 384523 379009 441088 441042 404378 383337 407642 375480 383073 383072 383186 420487 441239 414699 414689 414685 414702 414741 414692 388017 383269

Distribution pattern

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.