Timeline bucket
June 06, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
DRG Instruments GmbH

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

DRG Salivary Progesterone HS ELISA- IVD for the determination of progesterone in human saliva REF: SLV-5911 in vitro diagnostic

Z-2410-2019
Recall number
Z-2410-2019
Initiated
June 06, 2019
Classification
Class II
Status
Terminated
Recalling firm
DRG Instruments GmbH
Quantity
8 Kits

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Salivary Progesterone assay may observe a higher percentage of samples with low results. Saliva samples in the low measuring range of the kit {< 20 pg/ml) were assayed and producing a significantly lower result.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Salivary Progesterone assay may observe a higher percentage of samples with low results. Saliva samples in the low measuring range of the kit {< 20 pg/ml) were assayed and producing a significantly lower result.

Code information

Lot Number:315K029

Distribution pattern