device · product 1 of 1
ABL90 FLEX Analyzer REF 393090 UDI:05700693930909
- Recall number
- Z-2320-2019
- Initiated
- July 10, 2019
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Radiometer America Inc
- Quantity
- 13,042 units
App-derived interpretation
software security vulnerabilities with the firm's analyzer operating system, which may cause
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
Software Security; The action is being initiated because of software security vulnerabilities with the firm's analyzer operating system, which may cause the device to shutdown or reboot resulting in delayed medical treatment.
Code information
All lots with the operating system versions - RXPE -UIM2743 v1.3XXXXXX or RXPE-UIM910 v1.3.XXXXXX
Distribution pattern
US: AB, AL, AR, BC, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TX, VA, VT, WA, WI, WV OUS: Canada, Netherland, China, Czech Republic, Denmark, Germany, Spain, France, Hugary, , indiat Japan, South Korea, United Kingdom, Australia, Poland, South Africa, Singapore, Switzerland, Turkey, Norway, Austria, Italy, Sweden, Finland, Brazil, and Mexico