device · product 1 of 1
Serrato 9.5mm X 90mm Polyaxial Screw Catalog Number 482619590 Serrato 9.5mm X 100mm Polyaxial Screw Catalog Number 4826195100
Discrepant length; Lot B88751 90mm screws were manufactured with a 100mm length instead of the required 90mm length; lot B90243 100mm screws were manufactured with a 90mm length instead of the required 100mm length.
These labels are deterministic app interpretations, not FDA categories.
Discrepant length; Lot B88751 90mm screws were manufactured with a 100mm length instead of the required 90mm length; lot B90243 100mm screws were manufactured with a 90mm length instead of the required 100mm length.
Code information
Lot # B88751 UDI: (01)07613327351576(11)190125(10)B88751
Distribution pattern
US Nationwide Distribution - CA, FL, GA, MA, NY, PA, TX