Exact FDA event identifier

Event 83377

7 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 03, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Medical Solutions USA, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

7 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 7

SOMATOM Definition AS, Model Number 8098027 - Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. These CT systems can be used for low dose lunch cancer screening in high risk populations (as defined by professional medical societies).

Z-2268-2019
Recall number
Z-2268-2019
Initiated
July 03, 2019
Classification
Class II
Status
Terminated
Quantity
N/A

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software issues impacting systems utilizing syngo.CT software version VB10A which may result in

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort.

Code information

Serial Numbers: 98474 92280 92261 65370 66032 66649 65299 92298 64516 92335 92300 66062 64218 92313 92296 92254 64051 92342 92277 92307 92336 92315 92288 98473 65145 65493 92323 92321 92243

Distribution pattern

US Nationwide distribution.

device · product 2 of 7

SOMATOM Definition Edge, Model Number 10590000 - Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. These CT systems can be used for low dose lunch cancer screening in high risk populations (as defined by professional medical societies).

Z-2269-2019
Recall number
Z-2269-2019
Initiated
July 03, 2019
Classification
Class II
Status
Terminated
Quantity
N/A

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software issues impacting systems utilizing syngo.CT software version VB10A which may result in

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort.

Code information

Serial Numbers: 83904 83931 83834 83746 83811 83915 83801 83767 83880 83899 83852 83798 83837 83878 83832 83871 83903 83808 83763 75475 83732 83752 83382 83845 83846 83749 83847 83786 83842 83775 83796 94801 83768 83909 83875 83859 83928 83826 83754 83913 83896 83790 83824 83898 83810 83886 83758 83901 83914 83823

Distribution pattern

US Nationwide distribution.

device · product 3 of 7

SOMATOM Definition Flash, Model Number 10430603 - Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. These CT systems can be used for low dose lunch cancer screening in high risk populations (as defined by professional medical societies).

Z-2270-2019
Recall number
Z-2270-2019
Initiated
July 03, 2019
Classification
Class II
Status
Terminated
Quantity
N/A

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software issues impacting systems utilizing syngo.CT software version VB10A which may result in

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort.

Code information

Serial Numbers: 73008 98642

Distribution pattern

US Nationwide distribution.

device · product 4 of 7

SOMATOM Drive, Model Number 10431700 - Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. These CT systems can be used for low dose lunch cancer screening in high risk populations (as defined by professional medical societies).

Z-2271-2019
Recall number
Z-2271-2019
Initiated
July 03, 2019
Classification
Class II
Status
Terminated
Quantity
N/A

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software issues impacting systems utilizing syngo.CT software version VB10A which may result in

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort.

Code information

Serial Numbers: 105081 105152 105128 105123 105105 105150 105096 105047 105115 105086 105165 105121 105103 105141 105118 105109 105116 105080

Distribution pattern

US Nationwide distribution.

device · product 5 of 7

SOMATOM Confidence, Model Number 10590100 - Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. These CT systems can be used for low dose lunch cancer screening in high risk populations (as defined by professional medical societies).

Z-2272-2019
Recall number
Z-2272-2019
Initiated
July 03, 2019
Classification
Class II
Status
Terminated
Quantity
N/A

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software issues impacting systems utilizing syngo.CT software version VB10A which may result in

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort.

Code information

Serial Numbers: 100172 100091 100112 100143 100158 100191 100107 100157 100137 100190 100092 100129 100136 100189 100156 100117 100097 100099 100174 100073 100016 100182 100146 100170 100175 100090 100160 100087 100025 100165 100147 100046 100140 100085

Distribution pattern

US Nationwide distribution.

device · product 6 of 7

SOMATOM Edge Plus, Model Number 10267000 - Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. These CT systems can be used for low dose lunch cancer screening in high risk populations (as defined by professional medical societies).

Z-2273-2019
Recall number
Z-2273-2019
Initiated
July 03, 2019
Classification
Class II
Status
Terminated
Quantity
N/A

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software issues impacting systems utilizing syngo.CT software version VB10A which may result in

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort.

Code information

Serial Numbers: 122022 122020 122008 122036 122029 122044 122037 122009 122038 122035 122041 122011 122034 122004

Distribution pattern

US Nationwide distribution.

device · product 7 of 7

SOMATOM Force, Model Number 10742326 - Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. These CT systems can be used for low dose lunch cancer screening in high risk populations (as defined by professional medical societies).

Z-2274-2019
Recall number
Z-2274-2019
Initiated
July 03, 2019
Classification
Class II
Status
Terminated
Quantity
N/A

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software issues impacting systems utilizing syngo.CT software version VB10A which may result in

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort.

Code information

Serial Numbers: 75765 75949 94851 75984 76019 75789 75512 75770 75600 75994 76046 75768 75974 75867 75510 75535 75820 75754 75755 75993 75916 75971 75893 75944 75540 76027 75980 75571 75848 75953 75885 75500 75973 75850 75527 75728 75939 75923 83819 76057 75450 75760 75806 75808 75476 75588 75931 75900 75981 75906 75602 75964 75804 75837 75787 75743 76010 76041 75936 75959 75905 75942 75941 75805 75884 76006 75919 76003 75913 76050 76016 75524 75778 75501 75570 75769 75983 75849 76000 75999 75976 76008 75616 75750 76001

Distribution pattern

US Nationwide distribution.