Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 83426

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 22, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Zeus Scientific, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Dynex DSX Software-IVD ELISA Assays effected as follows: DSX Assay File Software: ZEUS vl B. burgdorferi lgG-lgM.asy ZEUS vl Borrelia VlsEl-pepClO lgG-lgM.asy ZEUS Parvovirus B19 lgG (9Z7701G).asy ZEUS Parvovirus B19 lgM (9Z7701M).asy ZEUS vl HSV-1 and 2 lgM.asy ZEUS vl Sm-RNP [CF] .asy ZEUS vl Sm-RNP [CV].asy ZEUS vl CMV lgG.asy ZEUS vl CMV lgM.asy ZEUS vl EBV-VCA lgG.asy ZEUS vl EBV-VCA lgM.asy ZEUS vl H. pylori lgG.asy ZEUS vl M. pneumoniae lgG.asy ZEUS vl M. pneumoniae lgM.asy ZEUS vl Measles lgG.asy ZEUS vl Mumps lgG .asy ZEUS vl T. gondii lgG .asy ZEUS vl T. gondii lgM.asy ZEUS vl VZV lgG.asy ZEUS vl Rubella lgG .asy ZEUS vl Rubella lgM.asy ZEUS vl MPO [CF].asy ZEUS vl MPO [CV].asy ZEUS vl EBNA-1 lgG.asy ZEUS vl.1 HSV-1 and 2 lgM.asy Product Usage: The DYNEX DSX Automated ELISA System is an open automated open system to perform ELISA immunoassay

Z-2494-2019
Recall number
Z-2494-2019
Initiated
July 22, 2019
Classification
Class II
Status
Terminated
Recalling firm
Zeus Scientific, Inc.
Quantity
28 Assays

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
programming error

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Assay files used on the open Dynex DSX Instrument to process IVD ELISA samples-programming error results in well H6 being skipped for the addition of the TMB substrate. Well H6 will generate a negative result regardless of whether the patient was negative or positive for that antibody. This could result in a potential false negative result for that assay for any patients tested in well H6.

Code information

Software versions: ZEUS vl B. burgdorferi lgG-lgM.asy ZEUS vl Borrelia VlsEl-pepClO lgG-lgM.asy ZEUS Parvovirus B19 lgG (9Z7701G).asy ZEUS Parvovirus B19 lgM (9Z7701M).asy ZEUS vl HSV-1 and 2 lgM.asy ZEUS vl Sm-RNP [CF] .asy ZEUS vl Sm-RNP [CV].asy ZEUS vl CMV lgG.asy ZEUS vl CMV lgM.asy ZEUS vl EBV-VCA lgG.asy ZEUS vl EBV-VCA lgM.asy ZEUS vl H. pylori lgG.asy ZEUS vl M. pneumoniae lgG.asy ZEUS vl M. pneumoniae lgM.asy ZEUS vl Measles lgG.asy ZEUS vl Mumps lgG .asy ZEUS vl T. gondii lgG .asy ZEUS vl T. gondii lgM.asy ZEUS vl VZV lgG.asy ZEUS vl Rubella lgG .asy ZEUS vl Rubella lgM.asy ZEUS vl MPO [CF].asy ZEUS vl MPO [CV].asy ZEUS vl EBNA-1 lgG.asy ZEUS vl.1 HSV-1 and 2 lgM.asy

Distribution pattern

US nationwide distribution in the states of CA, CT, FL, KY, IN, MA, MI, NY, NJ, NH, OH, PA, TX, VA.