Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 83462

4 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 10, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Medical Solutions USA, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

4 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 4

Lantis 6.1 Commander, Model # 5493072 - Product Usage: Lantis Treatstation is to allow the radiation therapist to deliver treatment to the patient using the mevatron and all available accessories. This entails selecting a patient, selecting today s treatment for that patient, setting up and delivering the treatment fields and recording the delivered treatment. Treatstation supports auto sequencing, a process of automatically downloading a group of fields or segments from the verification and record system to the control of the linear accelerator sequentially, without user intervention. In addition, Lantis Treatstation supports intensity modulation, a process of shaping, modifying and moving the beam around a target to maximize the dose at the target and minimize the dose to normal structures.

Z-2195-2019
Recall number
Z-2195-2019
Initiated
July 10, 2019
Classification
Class II
Status
Terminated
Quantity
2

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potentially affected by the Microsoft Windows Remote Desktop Protocol (RDP) vulnerability

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Other

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potentially affected by the Microsoft Windows Remote Desktop Protocol (RDP) vulnerability

Code information

Serial Number 93-1158 93-0803

Distribution pattern

State WV NY PA FL OK FL NJ NY IL WY NY GA NY PA PA AL NJ NY ***updated 8/14/19*** OH NC

device · product 2 of 4

Lantis 8.3 Commander, Model # 8148178 - Product Usage: Lantis Treatstation is to allow the radiation therapist to deliver treatment to the patient using the mevatron and all available accessories. This entails selecting a patient, selecting today s treatment for that patient, setting up and delivering the treatment fields and recording the delivered treatment. Treatstation supports auto sequencing, a process of automatically downloading a group of fields or segments from the verification and record system to the control of the linear accelerator sequentially, without user intervention. In addition, Lantis Treatstation supports intensity modulation, a process of shaping, modifying and moving the beam around a target to maximize the dose at the target and minimize the dose to normal structures.

Z-2196-2019
Recall number
Z-2196-2019
Initiated
July 10, 2019
Classification
Class II
Status
Terminated
Quantity
2

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potentially affected by the Microsoft Windows Remote Desktop Protocol (RDP) vulnerability

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Other

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potentially affected by the Microsoft Windows Remote Desktop Protocol (RDP) vulnerability

Code information

Serial Number 93-1296 93-1314 ***added 8/14/19*** 93-1531

Distribution pattern

State WV NY PA FL OK FL NJ NY IL WY NY GA NY PA PA AL NJ NY ***updated 8/14/19*** OH NC

device · product 3 of 4

Lantis Commander, Model # 4503178 - Product Usage: Lantis Treatstation is to allow the radiation therapist to deliver treatment to the patient using the mevatron and all available accessories. This entails selecting a patient, selecting today s treatment for that patient, setting up and delivering the treatment fields and recording the delivered treatment. Treatstation supports auto sequencing, a process of automatically downloading a group of fields or segments from the verification and record system to the control of the linear accelerator sequentially, without user intervention. In addition, Lantis Treatstation supports intensity modulation, a process of shaping, modifying and moving the beam around a target to maximize the dose at the target and minimize the dose to normal structures.

Z-2197-2019
Recall number
Z-2197-2019
Initiated
July 10, 2019
Classification
Class II
Status
Terminated
Quantity
15

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potentially affected by the Microsoft Windows Remote Desktop Protocol (RDP) vulnerability

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Other

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potentially affected by the Microsoft Windows Remote Desktop Protocol (RDP) vulnerability

Code information

Serial # 93-0759 93-0648 93-0662 93-1045 93-0574 93-0313 93-0501 93-0951 93-0667 93-0141 93-0786 93-0385 93-0799 93-0659 93-0861 ***added 8/14/19*** 93-0159 93-0878 93-0272 93-0596

Distribution pattern

State WV NY PA FL OK FL NJ NY IL WY NY GA NY PA PA AL NJ NY ***updated 8/14/19*** OH NC

device · product 4 of 4

Lantis system (pre-2003), Model #5835991 - Product Usage: Lantis Treatstation is to allow the radiation therapist to deliver treatment to the patient using the mevatron and all available accessories. This entails selecting a patient, selecting today s treatment for that patient, setting up and delivering the treatment fields and recording the delivered treatment. Treatstation supports auto sequencing, a process of automatically downloading a group of fields or segments from the verification and record system to the control of the linear accelerator sequentially, without user intervention. In addition, Lantis Treatstation supports intensity modulation, a process of shaping, modifying and moving the beam around a target to maximize the dose at the target and minimize the dose to normal structures.

Z-2198-2019
Recall number
Z-2198-2019
Initiated
July 10, 2019
Classification
Class II
Status
Terminated
Quantity
N/A

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potentially affected by the Microsoft Windows Remote Desktop Protocol (RDP) vulnerability

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Other

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potentially affected by the Microsoft Windows Remote Desktop Protocol (RDP) vulnerability

Code information

All serial numbers

Distribution pattern

State WV NY PA FL OK FL NJ NY IL WY NY GA NY PA PA AL NJ NY ***updated 8/14/19*** OH NC