device · product 1 of 1
AD-TECH Drill Sleeve Guides, used with Ad-Tech Placement Kits: (a) REF DSG-90-2.3N (b) REF DSG-6.3-090-2.4N
- Recall number
- Z-2544-2019
- Initiated
- August 27, 2019
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Ad-Tech Medical Instrument Corporation
- Quantity
- 315 units
App-derived interpretation
The inner diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
The inner diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
Code information
All serial numbers distributed between January 2014 and May 2019
Distribution pattern
NY, PA, WI, AZ, IL, CA, NH, GA, MA, MN, UT, MI, WA, FL, TX, OH, OR, PA, MD, and India, Korea, Taiwan, Canada, Hong Kong, Denmark, Russia, Singapore, Japan, Brazil. UK, Spain, Czech Republic, France